Physical and engineering sciences in medicine

Development and validation of a modular multi-purpose phantom for CT, PET/CT, and SPECT/CT quality control.

Abbas Masoumi, Mehrnoosh Karimipourfard, Seyed Mohammad Entezarmahdi, Mohammad Hossein Sadeghi, Sedigheh Sina

Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran. · Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran. mehrnoosh.karimipour@ucalgary.ca.

Computational

Key points

From the paper's conclusion
  • The observed agreement with available reference phantoms and consistency with established QC tolerance criteria support its potential use for routine CT, PET/CT, and SPECT/CT quality control applications.

Abstract

Commercial phantoms used for quality control (QC) of CT, PET/CT, and SPECT/CT systems are often designed for single tasks, requiring multiple devices to perform routine QC procedures. This study aimed to develop and validate a locally manufacturable, waterfillable, multi-purpose phantom capable of supporting several CT, PET/CT, and SPECT/CT quality control applications within a single modular platform. A PMMA-based phantom incorporating interchangeable inserts for CT dosimetry, spatial resolution, image registration, image uniformity, and contrast detectability assessment was designed and fabricated using thermal forming, CNC machining, and solvent bonding. Experimental evaluation was performed on clinical CT, PET/CT, and SPECT/CT systems using comparisons with standard/reference phantoms where available and established QC tolerance criteria. Manufacturer performance data were used only as secondary contextual benchmarks when an independent reference phantom was unavailable. CT dosimetry measurements demonstrated good agreement with a standard CTDI phantom, with CTDIV differences of 4.5% and 2.3% for head and body configurations, respectively. PET spatial resolution measurements yielded FWHM values of 3.72-3.91 mm, while PET/CT registration errors remained below one pixel (0.651 and 0.118 pixels). PET uniformity measurements showed values of 1.32-1.35%, with less than 10% deviation from a reference ACR phantom. SPECT spatial resolution measurements ranged from 8.6 to 10.8 mm; manufacturer catalogue values were used only as secondary contextual benchmarks for these measurements because an independent reference phantom was unavailable. SPECT/CT registration errors were 1.8 mm and 0.67 mm, both below accepted tolerance limits. SPECT uniformity measurements yielded integral and differential uniformity values of 1.5%, 0.9%, and 0.8%, respectively. In contrast detectability experiments, all inserts were visualized at 0.7 mCi, whereas the smallest insert became indistinguishable from the background at 0.3 mCi. The proposed phantom provides a practical and cost-effective solution for multi-modality quality control by integrating CT dosimetry, spatial resolution assessment, image registration evaluation, image uniformity analysis, and contrast detectability testing within a single reusable platform. The observed agreement with available reference phantoms and consistency with established QC tolerance criteria support its potential use for routine CT, PET/CT, and SPECT/CT quality control applications.

Read the paper