The impact of data extraction percentage and deep learning-based reconstruction on image quality in gated PET/computed tomography.
Hiroki Nosaka, Masaya Suda, Noriaki Miyaji, Hiraku Fuse, Kenji Yasue, Norikazu Koori, Shin Miyakawa, Masato Takahashi,
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Inashiki, Ibaraki. · Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima.
Key points
From the paper's conclusion- The combination of DDG and AiCE-i maintained stable image quality across the respiratory conditions evaluated. Under the conditions of this phantom study, a %count range of 30-40% provided a favorable balance between image noise and the effects of respiratory motion for standard 120-s acquisitions.
Abstract
BACKGROUND: Respiratory motion artifacts degrade PET/computed tomography (PET/CT) image quality. Data-driven gated (DDG) PET/CT addresses this issue by extracting respiratory signals directly from PET data, eliminating the need for external monitoring devices. This study investigated the effects of data extraction percentage (%count) and deep learning-based reconstruction [Advanced Intelligent Clear-IQ Engine-integrated (AiCE-i)] on image quality in DDG-PET under different respiratory conditions using a phantom model. METHODS: A body phantom containing six spheres (10-37 mm) was imaged using a silicon photomultiplier-based PET/CT system. Four respiratory waveforms (no-motion, sinusoidal, representative patient, and baseline shift) and four %count levels (20, 30, 40, and 50%) were evaluated using AiCE-i reconstruction. Image quality was assessed using background variability (N10 mm), percentage contrast (QH,10 mm), contrast-to-noise ratio (QH,10 mm/N10 mm), and recovery coefficient. RESULTS: Increasing %count consistently reduced N10 mm across all respiratory waveforms. QH,10 mm and QH,10 mm/N10 mm generally increased with increasing %count; however, at the 50% threshold, significant reductions were observed in the baseline shift and sinusoidal waveforms compared with the no-motion condition. Recovery coefficient analysis demonstrated the partial volume effect in smaller spheres and showed that quantitative performance was maintained across the evaluated gating conditions. CONCLUSION: The combination of DDG and AiCE-i maintained stable image quality across the respiratory conditions evaluated. Under the conditions of this phantom study, a %count range of 30-40% provided a favorable balance between image noise and the effects of respiratory motion for standard 120-s acquisitions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reconstruction-dependent differences in brain [18F]FDG PET: a large-scale same-acquisition comparison of Q.Clear and VUE Point FX.Annals of nuclear medicine · 2026
- Count-aware diffusion with autoregressive inference for low-count PET reconstruction enhancement.Medical image analysis · 2026
- Artificial Intelligence Across the PET Reconstruction Pipeline: An Update.AJR. American journal of roentgenology · 2026
- New reconstruction algorithm for stack artifact reduction in breathing-adaptive four-dimensional computed tomography.Physics and imaging in radiation oncology · 2026
- Qualitative and quantitative assessment of deep learning-based raw data motion correction in atrial fibrillation: benefit in high heart rate and motion-prone segmentsEuropean Heart Journal · 2026
- A deep learning approach for noise suppression in cardiac-gated single-photon emission computed tomography studies.Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2026