New reconstruction algorithm for stack artifact reduction in breathing-adaptive four-dimensional computed tomography.
Juliane Szkitsak, Maximilian Heider, Niklas Andre Lackner, Charlotte Frei, Allison Lamrani, Christopher Dürrbeck, Annika Oleszczuk, Mushawar Ahmad,
Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany. · Siemens Healthineers AG, 91301 Forchheim, Germany.
Key points
From the paper's conclusion- This study demonstrated a statistically significant reduction of stack artifacts, resulting in improved anatomical correctness when ZeeFree was applied to 4DCTs. The algorithm was not observed to introduce new artifacts on already artifact-free images.
Abstract
BACKGROUND AND PURPOSE: Even with new adaptive four-dimensional computed tomography (4DCT) techniques, artifacts can still occur with strong irregular breathing amplitudes. A new image reconstruction algorithm, currently available for cardiac CT, was tested on 4DCT data for its performance to reduce stack artifacts and thus improve anatomical correctness. MATERIALS AND METHODS: The 'ZeeFree' algorithm employs a deformable image registration during reconstruction to reduce remaining stack artifacts. Its performance was evaluated using data of 20 irregular breathing ('challenging group') and 10 randomly selected ('verification group') patients with tumors in the thorax. Using a 5-score-scale, image quality in terms of stack artifacts was rated by three physicians and three physicists. Statistical significance was assessed using the Wilcoxon signed-rank test for the mean ratings and the Chi-square test for anatomical correctness. RESULTS: Expert scoring revealed that 88.3% of the amplitude-based ZeeFree reconstructions showed (almost) no artifacts (rated with ≥ 4) compared to 66.8% of the standard reconstruction. With phase-based binning, 70.0% of the ZeeFree and 23.3% of standard were rated with a score ≥ 4. Mean scores were 4.1 and 4.0 for amplitude- and phase-based binning with ZeeFree and 3.4 and 2.5 for standard reconstruction, respectively. Differences were significant (p < 0.001). 91.1% (ZeeFree) and 41.1% (standard) were rated as anatomically correct. Again, differences were significant (p < 0.001). CONCLUSIONS: This study demonstrated a statistically significant reduction of stack artifacts, resulting in improved anatomical correctness when ZeeFree was applied to 4DCTs. The algorithm was not observed to introduce new artifacts on already artifact-free images.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction.Physics and imaging in radiation oncology · 2026
- A clinical evaluation of spiral 4DCT and intelligent 4DCT sequence scanning.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Improving image quality by reducing breathing artifacts in CT of patients early after coronary artery bypass graft surgery.The international journal of cardiovascular imaging · 2026
- Improving depth-resolution, in-plane contrast, and reducing non-uniformity artifacts for wide-angle DBT.ArXiv · 2026
- Improving depth-resolution, in-plane contrast, and reducing nonuniformity artifacts for wide-angle digital breast tomosynthesis.Journal of medical imaging (Bellingham, Wash.) · 2026
- Task-based image quality and dosimetric validation of HyperSight CBCT for adaptive radiotherapy.Journal of applied clinical medical physics · 2026