Diagnostic accuracy of prehospital capillary high-sensitivity cardiac troponin I in patients with acute chest pain: the prospective multicenter VALIDATE study.
Jérémy Guenezan, Clementine Cassard, Anthony Chauvin, Samir Bouhadiba, Raphaël Couvreur, Pierre Dupuis, Yohann Foucher, Fanny Hamard,
Emergency Department, University Hospital of Poitiers, Faculty of Medicine, University of Poitiers, Poitiers. · Emergency Department and Prehospital Emergency Medical Service (SAMU/SMUR), GHU Sorbonne Université, Pitié-Salpêtrière Hospital, Assistance Publique - Hôpitaux de Paris (AP-HP).
Key points
From the paper's conclusion- Prehospital capillary hs-cTnI showed good diagnostic performance for NSTEMI, although feasibility in routine practice was challenged by unavailable or uninterpretable measurements. Further external validation is warranted before implementation in prehospital diagnostic pathways.
Abstract
BACKGROUND: High-sensitivity cardiac troponin I (hs-cTnI) plays a central role in the diagnosis and management of non-ST-elevation myocardial infarction (NSTEMI), including rule-out and rule-in strategies in the emergency department. Extending hs-cTnI testing to the prehospital setting could accelerate patient orientation and clinical decision-making, but the diagnostic performance of prehospital capillary hs-cTnI remains insufficiently studied. OBJECTIVE: The study aimed to evaluate the diagnostic performance of prehospital capillary hs-cTnI for the diagnosis of NSTEMI at hospital discharge. DESIGN: The study is a prospective multicenter diagnostic accuracy study with blinded index test results. SETTINGS AND PARTICIPANTS: Four French emergency medical service systems with 10 mobile intensive care units. Adult patients presenting with acute chest pain were included. INDEX TEST: Prehospital hs-cTnI measurement using a point-of-care device (Atellica VTLi) on capillary samples. Treating clinicians were blinded to hs-cTnI results. OUTCOME MEASURES AND ANALYSIS: The primary outcome was the diagnostic performance of capillary hs-cTnI for the diagnosis of NSTEMI at hospital discharge. Diagnostic performance was assessed using the area under the receiver operating characteristic curve with 95% confidence intervals (CIs). Clinically relevant thresholds were defined a priori based on predefined safety criteria. MAIN RESULTS: Among 659 patients included in the primary analysis, 71 (10.8%) had NSTEMI. Prehospital capillary hs-cTnI demonstrated good diagnostic performance [area under the curve (AUC) = 0.83, 95% CI = 0.78-0.88]. A rule-out threshold of <2.2 ng/L, evaluated according to prespecified performance criteria, identified 7.1% of patients, with a negative predictive value of 1.00 (95% CI = 0.92-1.00). A rule-in threshold of >43.6 ng/L identified 6.5% of patients, with a positive predictive value of 0.65 (95% CI = 0.51-0.79). In the predefined sensitivity analysis, venous hs-cTnI showed similar diagnostic performance (AUC = 0.84, 95% CI = 0.79-0.89). CONCLUSION: Prehospital capillary hs-cTnI showed good diagnostic performance for NSTEMI, although feasibility in routine practice was challenged by unavailable or uninterpretable measurements. Further external validation is warranted before implementation in prehospital diagnostic pathways.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic accuracy of high-sensitivity point-of-care troponin assays in emergency department for patients presenting with chest pain: a systematic review.European journal of emergency medicine : official journal of the European Society for Emergency Medicine · 2026
- Cost-effectiveness of prehospital high-sensitivity troponin i testing versus standard emergency department care in patients with suspected acute coronary syndrome in France.Health economics review · 2026
- Diagnosis of myocardial infarction in early presenters using 1-hour delta of high sensitive Troponin I: a pilot study.Clinics (Sao Paulo, Brazil) · 2026
- Real-world analytical and diagnostic evaluation of the Abbott i-STAT Alinity point-of-care high-sensitivity troponin I assay in a UK emergency department.Annals of clinical biochemistry · 2026
- ICare-FASTER (Improving Care by Faster Risk Stratification Through Use of High-Sensitivity Point-of-Care Troponin in Patients Presenting With Possible Acute Coronary Syndrome in the Emergency Department): A Stepped-Wedge Cluster-Randomized Trial.Circulation · 2026
- Comparison of Troponin in a Prehospital Blood Sample and Risk Scores.Diagnostics (Basel, Switzerland) · 2026