InSafeLock endopin versus conventional distal locking screws for femoral shaft fractures: A comparative laboratory biomechanical study.
Seçkin Özcan, Hakan Yurten, Zekeriya Ersin Çelen
Department of Orthopaedics and Traumatology, Yalova Training and Research Hospital, Yalova University Faculty of Medicine, Yalova, Turkey. · Department of Orthopaedics and Traumatology, Ordu Meditech Private Hospital, Ordu, Turkey.
Key points
From the paper's conclusion- While the InSafeLock endopin offers ease in distal locking, it exhibits a significantly lower failure load under high axial compression forces than conventional screws.
Abstract
The InSafeLock femoral nail is a recently developed implant that uses endopin instead of screws in its distal locking mechanism. This study aimed to compare the biomechanical features of the distal locking mechanisms of InSafeLock and conventional nails. Twenty-four artificial femoral bones were used for mechanical tests. InSafeLock nails with endopin (Group 1) and conventional nails with distal locking screws (Group 2) were inserted into the artificial bones. The biomechanical features were evaluated using axial compression and rotational performance tests. There was no statistically significant difference between Group 1 and Group 2 in the axial compression displacement tests (P = .683). No visual damage was observed on the artificial bones or intramedullary nails after axial compression fatigue tests in either group. During axial compression failure testing performed after fatigue testing, the mean failure load of Group 1 was significantly lower than that of Group 2 (1132.5 ± 15.5 N vs 11,965.5 ± 9.5 N, respectively; P = .002). Rotational performance tests showed that the torsional yield strength of the groups was not statistically different (P = .333). While the InSafeLock endopin offers ease in distal locking, it exhibits a significantly lower failure load under high axial compression forces than conventional screws.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Internal-locking intramedullary nails (talons, claws, wings, wedge-locks, endopins): a mechanics-guided scoping review of designs, failure signatures, and technique imperatives.BMC musculoskeletal disorders · 2026
- Increased screw backout with non–inlay design retrograde femoral nailsEuropean Journal of Orthopaedic Surgery & Traumatology · 2026
- Slot-targeted longitudinal unicortical femorotomy for the extraction of an incarcerated femoral intramedullary nail during hip arthroplasty: a case reportFrontiers in Surgery · 2026
- Intramedullary Femoral Nailing-Blade Versus Screw: A Risk Analysis of Implant-Related Complications in Trochanteric Femoral Fractures.Journal of clinical medicine · 2026
- The impact of subsidence on hip function: A comparative study between Proximal Femoral Nail Anti-rotation-2 and Halifax Femoral Nail in unstable trochanteric fractures.SICOT-J · 2026
- Investigation of the biomechanically optimal length of intramedullary nails in unstable intertrochanteric femur fractures 31A2.2: A finite element analysis study.Injury · 2026