Performance evaluation of electrosurgical pencils: insights into surgical smoke mitigation and thermal damage reduction.
Kihyun Sang, Yoon Kong, Ja Kyung Lee, Hee Young Na, Hyeong Won Yu, June Young Choi
Department of Surgery, Seoul National University Bundang Hospital, Seongnam-si, Gyeonggi-do, Republic of Korea. · Department of Pathology, Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
Key points
From the paper's conclusion- Duoblades, especially those with suction, consistently produced the lowest level of surgical smoke, reducing particle counts by up to 96.1% compared to the uncoated Bovie. It also achieved the greatest reduction in blood loss (93.3% less than that of the scalpel) and resulted in the narrowest scar width among the ESPs. The wound scar width and densities of CD3-positive lymphocytes and myofibroblasts were reduced more rapidly in Duoblade-incised wounds than in those with other ESPs.
Abstract
BACKGROUND: Electrosurgical pencils (ESPs) are essential in modern surgery; however, concerns remain regarding their exposure to surgical smoke and thermal tissue damage. This study quantitatively evaluated four ESPs with a particular focus on two critical aspects: surgical smoke generation and minimization of thermal damage to the surrounding tissues. METHODS: Full-thickness incisions were made on the porcine dorsal skin using six instruments: scalpel, uncoated Bovie, coated Bovie, PlasmaBlade, and Duoblade (with and without suction). Surgical smoke was measured by particle counting, blood loss was quantified, and histological analyses (hematoxylin and eosin, Masson's trichrome, CD3, and α-smooth muscle actin staining) were performed to assess the wound scar width and densities of CD3-positive lymphocytes and myofibroblasts. RESULTS: Duoblades, especially those with suction, consistently produced the lowest level of surgical smoke, reducing particle counts by up to 96.1% compared to the uncoated Bovie. It also achieved the greatest reduction in blood loss (93.3% less than that of the scalpel) and resulted in the narrowest scar width among the ESPs. The wound scar width and densities of CD3-positive lymphocytes and myofibroblasts were reduced more rapidly in Duoblade-incised wounds than in those with other ESPs. DISCUSSION: Device design and energy delivery significantly affect the surgical smoke and tissue outcomes. Duoblade dielectric heating and integrated suction provide superior environmental safety and wound healing, while maintaining excellent hemostasis. The Duoblade has a suction function that reduces the surgical smoke.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating surgical smoke in laser-assisted laparoscopic partial nephrectomy with a novel handle: anin Vitrostudy.The Canadian journal of urology · 2026
- Management of aerosol generating particles with a novel self-retaining intraoral suction device.International journal of pediatric otorhinolaryngology · 2026
- Electronic Bandage With High‐Resolution Interdigital Electrode Enhanced Pulsed Electric Field for Accelerated Wound HealingEcoMat · 2026
- Bio-inspired electrosurgical knife with lubricant-assisted porous-groove surface and micro-serrated edge for reducing tissue adhesion and thermal injury.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
- Biodegradable piezoelectric suture for real-time biomechanical sensing and therapeutic intervention.Science bulletin · 2026
- Invisible shield: sprayable supramolecular antimicrobial microscale films for preventing wound and medical device infections.npj biomedical innovations · 2026