Dynamic Noninvasive Assessment of Intracranial Pressure, Cerebral Perfusion Pressure, and Autoregulation.
Ihsane Olakorede, Stefan Yu Bögli, Marek Czosnyka, Peter Smielewski
Department of Clinical Neurosciences, Division of Neurosurgery, Brain Physics Laboratory, University of Cambridge, Cambridge, UK.
Key points
From the paper's conclusion- Nonoverlapping 5-minute segments from 96 simultaneous intracranial pressure (ICP) and TCD recordings in 38 TBI patients were analyzed. Eight blood velocity-based models were used to derive noninvasive estimates of ICP, CPP, pressure reactivity index (PRx), and mean flow index (Mx). Agreement was assessed using correlation and Bland-Altman analyses, whereas dynamic performance was evaluated using first-order differences and linear mixed-effects modeling.
- Noninvasive estimates of CPP and Mx demonstrated stronger dynamic agreement with their reference measures than noninvasive estimates of ICP and PRx. Selected models showed robust discrimination of low CPP and impaired Mx (area under the curve up to 0.96 and 0.89). Conversely, reproduction of ICP dynamics and PRx was moderate, with significant interpatient and intrapatient variability.
- Evaluation of noninvasive ICP monitoring should extend beyond static accuracy. TCD-based approaches may support noninvasive monitoring of CPP and Mx, but current methods remain insufficient for reliable tracking of ICP and PRx dynamics.
Abstract
BACKGROUND: Intracranial pressure (ICP) monitoring is critical for managing traumatic brain injury, yet invasive methods carry risks and require specialized settings. Noninvasive transcranial Doppler (TCD)-based methods have been proposed to estimate ICP and cerebral perfusion pressure (CPP), although validation studies predominantly assess static agreement. Clinical monitoring depends on accurate reproduction of dynamic changes. This study assesses the ability of noninvasive methods to reproduce temporal variations in ICP, CPP, and autoregulatory indices. METHODS: Nonoverlapping 5-minute segments from 96 simultaneous ICP and TCD recordings in 38 TBI patients were analyzed. Eight blood velocity-based models were used to derive noninvasive estimates of ICP, CPP, pressure reactivity index (PRx), and mean flow index (Mx). Agreement was assessed using correlation and Bland-Altman analyses, whereas dynamic performance was evaluated using first-order differences and linear mixed-effects modeling. Discriminative ability for intracranial hypertension (ICP>22 mm Hg), low CPP (<60 mm Hg), and impaired autoregulation (PRx>0.3; Mx>0.4) was examined using receiver operating characteristic and precision-recall analyses. RESULTS: Noninvasive estimates of CPP and Mx demonstrated stronger dynamic agreement with their reference measures than noninvasive estimates of ICP and PRx. Selected models showed robust discrimination of low CPP and impaired Mx (area under the curve up to 0.96 and 0.89). Conversely, reproduction of ICP dynamics and PRx was moderate, with significant interpatient and intrapatient variability. Dynamic analyses revealed discrepancies missed by static agreement metrics. CONCLUSIONS: Evaluation of noninvasive ICP monitoring should extend beyond static accuracy. TCD-based approaches may support noninvasive monitoring of CPP and Mx, but current methods remain insufficient for reliable tracking of ICP and PRx dynamics.
Read the paper
Discussion · 1
Science Skim's AI key points, plus readers' public thoughts
- Nonoverlapping 5-minute segments from 96 simultaneous intracranial pressure (ICP) and TCD recordings in 38 TBI patients were analyzed. Eight blood velocity-based models were used to derive noninvasive estimates of ICP, CPP, pressure reactivity index (PRx), and mean flow index (Mx). Agreement was assessed using correlation and Bland-Altman analyses, whereas dynamic performance was evaluated using first-order differences and linear mixed-effects modeling.
- Noninvasive estimates of CPP and Mx demonstrated stronger dynamic agreement with their reference measures than noninvasive estimates of ICP and PRx. Selected models showed robust discrimination of low CPP and impaired Mx (area under the curve up to 0.96 and 0.89). Conversely, reproduction of ICP dynamics and PRx was moderate, with significant interpatient and intrapatient variability.
- Evaluation of noninvasive ICP monitoring should extend beyond static accuracy. TCD-based approaches may support noninvasive monitoring of CPP and Mx, but current methods remain insufficient for reliable tracking of ICP and PRx dynamics.
Readers' thoughts
No reader thoughts yet.
Related papers
- Toward comprehensive multimodal neuromonitoring of surrogates of cerebral autoregulation in traumatic brain injury: insights from a hyperventilation protocol.Neurophotonics · 2026
- Intracranial Pressure Estimation Using B4C Device: A Systematic Review on the Validation Studies.Neurocritical care · 2026
- Noninvasive Intracranial Pressure Estimation Using Transcranial Doppler to Rule Out Intracranial Hypertension in Children Undergoing Lumbar Puncture: A Feasibility Study.Pediatric neurology · 2026
- Noninvasive Intracranial Pressure Monitoring in Pediatric Traumatic Brain Injury: Systematic Review of Transcranial Doppler and Optic Nerve Sheath Diameter With Identification of an Evidence Gap in Pupillometry.Pediatric neurology · 2026
- Association Between Individualized Intracranial Pressure Thresholds and Long-Term Outcomes in Moderate-to-Severe TBI: A CAHR-TBI Multicenter Study.Journal of neurotrauma · 2026
- Intracranial Pressure Variability and Low Cerebral Perfusion Pressure Burden as Prognostic Markers Beyond Mean Pressure in Neurocritically Ill Patients.Journal of clinical medicine · 2026