Development of a Rapid Automated Point-of-Care Test for Mycobacterium tuberculosis Detection From Tongue Swabs and Sputum Specimens on the DASH® Rapid PCR System.
Matthew A Butzler, Jennifer L Reed, Alaina M Olson, Rachel C Wood, Gerard A Cangelosi, Angelique K Luabeya, Mark Hatherill, Arthur M Chiwaya,
Center for Innovation in Global Health Technologies, Robert J. Havey Institute for Global Health, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA. · Department of Environmental and Occupational Health Sciences, University of Washington, Seattle, Washington, USA.
Key points
From the paper's conclusion- The DASH MTB's assay demonstrated high sensitivity and specificity, rapid time to result, and compatibility with environments that lack air conditioning and consistent electricity make this assay suitable for diverse clinical settings.
Abstract
BACKGROUND: Mycobacterium tuberculosis (MTB) disease is a global health threat with most cases occurring in low-and middle-income countries (LMIC) with limited healthcare infrastructure. Near-point-of-care testing deployed at peripheral clinical settings is needed to start treatment earlier and thereby improve treatment outcomes. METHODS: We developed a rapid MTB detection assay for the DASH® Rapid PCR System with cartridge-based sample prep and dual target qPCR multicopy MTB insertion sequences IS6110 and IS1081 amplification. The limit of detection and analytical specificity of the RUO (research use only) DASH MTB assay were evaluated. Two clinical evaluations were performed using 110 (70 positive and 40 negative) de-identified blinded tongue swab specimens and 100 (49 positive and 51 negative) de-identified blinded sputum specimens from South African symptomatic clinic attendees. RESULTS: The DASH assay demonstrated a limit of detection of 2.5 MTB cells/swab with no detection of 10 non-tuberculosis Mycobacterium strains. The clinical sensitivity of tongue swab specimens was 91%, and the specificity was 100% as compared to Xpert Ultra results on paired sputum specimens. The overall clinical sensitivity of sputum specimens was 96% (100% for smear-positive and 88% for smear-negative samples) and the specificity of 88% when compared to sputum culture. The DASH test is compatible with peripheral LMIC settings via external battery operation and cartridge stability at 45°C for one year. CONCLUSIONS: The DASH MTB's assay demonstrated high sensitivity and specificity, rapid time to result, and compatibility with environments that lack air conditioning and consistent electricity make this assay suitable for diverse clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic accuracy of a novel point-of-care tongue swab assay for pulmonary tuberculosis: a multicentre prospective studyClinical Microbiology and Infection · 2025
- Pulmonary Tuberculosis Detection with MiniDock MTB Using Swab SamplesNew England Journal of Medicine · 2026
- Diagnostic Performance of the Pluslife MiniDock MTB and Molbio MTB Ultima Assays to Detect Tuberculosis From Tongue and Sputum Swabs Among Outpatients and in Active Case Finding in CameroonClinical Infectious Diseases · 2025
- Diagnostic Accuracy and Usability of the Pluslife Mycobacterium tuberculosis Assay on Tongue and Sputum Swabs in Symptomatic and Asymptomatic Adults in South Africa.Diagnostics (Basel, Switzerland) · 2026
- Diagnostic accuracy of TB-LAMP assay on tongue swabs for pulmonary tuberculosis.Microbiology spectrum · 2026
- Diagnostic accuracy of swab-based molecular tests for tuberculosis using near-point-of-care platforms: a multi-country evaluationEBioMedicine · 2025