Performance of intraoperative artificial intelligence-assisted label-free stimulated Raman histology in recurrent brain tumors.
Gina Fürtjes, Abdulkader Al Shughri, Thomas Stehle, David Reinecke, Anna-Katharina Meißner, Nina Jung, Zeynep Mercan, Nicolas Neidert,
Department of General Neurosurgery, Centre of Neurosurgery, University Hospital Cologne, Cologne, Germany. · Institute for Neuropathology, University Hospital Cologne, University of Cologne, Cologne, Germany.
Key points
From the paper's conclusion- Reliable SRH-based diagnosis in recurrent tumors requires validated image-quality control measures and model adaptation to recurrent disease.
Abstract
BACKGROUND: Therapy-induced tissue alterations complicate intraoperative diagnosis of recurrent central nervous system (CNS) tumors. We evaluated stimulated Raman histology (SRH) combined with convolutional neural networks (CNNs) for rapid tumor detection and classification. METHODS: In this bicentric observational study, fresh tissue from patients undergoing resection or stereotactic biopsy for suspected recurrent CNS tumors was imaged with fiber-laser-based SRH. CNN1 performed binary tumor/non-tumor prediction and CNN2 histotype classification. CNN1 was evaluated against blinded neuropathologist SRH consensus and CNN2 against the final integrated histopathological diagnosis. RESULTS: Of 382 acquired SRH images, 262 images from 143 patients were included. CNN1 achieved an AUC of 0.975 (95% CI, 0.95-1.00) for tumor detection. An exploratory cohort-derived tumor probability threshold of ≥59 increased specificity from 70.7% to 92.0% while maintaining 95.2% sensitivity and identifying 71.4% of radiation necrosis samples. Among 212 images in treatment subgroup analysis, CNN1 accuracy did not differ significantly across prior-treatment groups (overall 90.1%; P = .080). CNN2 accuracy declined from 61.8% after surgery alone to 39.0% after surgery, radiotherapy, and medical treatment (χ2 = 6.04, P = .049). Multimodal therapy was independently associated with lower odds of correct CNN2 prediction (adjusted OR, 0.10 [95% CI, 0.02-0.39]; P = .001 and 0.04 [95% CI, 0.01-0.20]; P < .001). 120/382 images (31.4%) were of insufficient diagnostic quality; application of the quality control threshold would have rejected 84/120 (70.0%). CONCLUSIONS: CNN1 maintained high tumor/non-tumor discrimination across treatment subgroups when image quality was sufficient, whereas CNN2 entity-classification performance declined with increasing treatment intensity. Reliable SRH-based diagnosis in recurrent tumors requires validated image-quality control measures and model adaptation to recurrent disease.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Artificial intelligence-based tumor detection using stimulated Raman histology in patients undergoing neurosurgical biopsies.Neurosurgical focus · 2026
- PATH-10 SIGNAL: A Scalable, Real-World Model for Rapid Intraoperative Molecular Classification of Gliomas Using Stimulated Raman HistologyNeuro-Oncology · 2026
- SURG-15 Fast, intraoperative, single-cell tumor infiltration detection in brain tumorsNeuro-Oncology · 2026
- The Intraoperative Utility of Raman Spectroscopy for Neurosurgical OncologyCancers · 2025
- Evaluation of Stimulated Raman Histology as a Novel Method for Pathologic Examination of Urothelial Carcinoma With Unprocessed, Fresh Bladder Cancer Tissue.Technology in cancer research & treatment · 2026
- IMG-08 Deciphering the FLAIR Signal: Stimulated Raman Histology Reveals the Limited Ability of Radiographic Characteristics to Comprehensively Define Tumor Infiltration in GlioblastomaNeuro-Oncology · 2026