EP03.19 Procedural comparison between classical and digital intraoperative rapid section diagnosis with multiphoton microscopy in neurosurgery
L Dührsen, M Middelkamp, M Dottermusch, J Matschke, F L Ricklefs, M Mohme, M Glatzel, J E Neumann
Key points
From the paper's conclusion- These findings suggest that DRSD can provide reliable histological feedback in a substantially shorter timeframe while retaining tissue for potential subsequent molecular analyses.
Abstract
Background Intraoperative neuropathology is essential for guiding surgical decisions and optimizing sampling strategies in neuro-oncological procedures. Conventional tissue-destructive rapid section diagnosis (CRSD) involves time-consuming tissue processing steps, which may delay critical intraoperative feedback. Digital rapid section diagnosis (DRSD) via multiphoton microscopy (MPM) retains the tissue and can represent a promising alternative, potentially offering expedited and more efficient diagnostics. This study aims to compare a novel MPM-based digital workflow to the established CRSD approach, assessing its potential benefits for intraoperative neuropathology in neuro-oncology. Material and Methods A monocentric cohort study was conducted at the University Medical Center Hamburg-Eppendorf, encompassing 62 neurosurgical tumor samples of 51 patients. Parallel analyses were performed using CRSD (H&E stained cryosections) and DRSD (Digital images from multiphoton microscopy (MPM) (Histolution GmbH, Lübeck, Germany)). Each CRSD and DRSD were evaluated independently by two board certified neuropathologists to assess interrater variability. Key parameters evaluated included quality of histomorphological image data, diagnostic turnaround and diagnostic accuracy (ground truth: postoperative diagnosis of Formalin fixed and Paraffin embedded material after full neuropathological workup). Results Our cohort comprised meningiomas (n = 11), diverse gliomas (n = 15), metastases (n = 2), pituitary neuroendocrine tumors (n = 6), mixed other tumor entities (n = 12) and non-neoplastic / reactive tissue (n = 5). All samples underwent parallel intraoperative CRSD and DRSD evaluation. For DRSD the scan time per sample was 2:34 min +/- 1:13 min. Time to diagnosis was signifcantly reduced comparing DRSD and CRSD (10 min vs. 19 min, p < 0.001). Qualitatively DRSD could display classical histomorphology features of brain tumors that could be evaluated by neuropathologists without specific training. Neuropathological evaluations based on CRSD and DRSD could reliably depict brain tumor entities with comparable accuracy and interrater variability was similar for CRSD and DRSD. These findings suggest that DRSD can provide reliable histological feedback in a substantially shorter timeframe while retaining tissue for potential subsequent molecular analyses. Conclusion MPM-enabled DRSD offers a tissue-friendly, efficient workflow by imaging intact, unfixed or fixated specimens and eliminating the need for tissue freezing or sectioning. Rapid, high-resolution digital histology images enable telepathology and real-time collaboration, simplify logistics, and preserve tissue for molecular testing. DRSD advances intraoperative neuropathology, supporting flexible sampling and timely decisions in neuro-oncology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- P07.01.A Near-real-time virtual histology for intraoperative neuro-oncology using multiphoton microscopy in 100 neurosurgical specimensNeuro-Oncology · 2026
- Feasibility of Real-Time Cellular Imaging Using a Miniaturized Epifluorescence Widefield Microscope for Intraoperative Margin Assessment in Intracranial Tumor Resection: A Pilot Study.World neurosurgery · 2026
- Double Staining Immunohistochemistry and Digital Pathology: Moving Towards Standardization of the Proliferative Index Evaluation in MeningiomasJournal of Personalized Medicine · 2026
- Shining a light on brain tumors: a review of Raman spectroscopy in the operating room.Journal of biomedical optics · 2026
- Label-free dynamic cell imaging of fresh ex vivo spinal meningiomas: a nine-case optical-pathological correlation study with specimen-level H&E references.Diagnostic pathology · 2026
- Lipid Metabolism-Related Rapid Pathological Diagnosis and Staging of Breast Tissue Sections Using Two-Photon Fluorescence Imaging Assisted with a Carbonized Polymer Dot Probe.Chemical & biomedical imaging · 2026