Neuro-Oncology

SURG-38 Canonical stereotactic trajectories and technical considerations for image-guided LITT

Chen Li, Shruti Gupta, Kylie Biegler, Chris Osswald, Lyubomir Zagorchev, Hsien-Chung Chen, Andrew Venteicher, Marcus Weimar,

Observational

Key points

From the paper's conclusion
  • We identified five reproducible, anatomy‑based canonical trajectories for ClearPoint LITT procedures performed within the MRI bore with the ClearPoint Prism system.

Abstract

ClearPoint MRI‑guided stereotactic procedures enable precise, real‑time targeting, yet optimal surgical corridors remain highly operator‑dependent and lack standardized guidance for minimally invasive LITT. Given the variation in tumor locations and the confined physical workspace of the MRI bore during intervention, where instrument maneuverability and patient positioning are inherently restricted, the need for efficient, pre‑defined workflows becomes even more critical. We analyzed 107 consecutive MRI‑guided ClearPoint stereotactic LITT procedures (2017-24), extracting demographic, anatomical, and trajectory‑specific variables for each case. All image volumes were transformed to a normalized image space via rigid image registration to the MNI152 brain atlas. A modified K-Means unsupervised algorithm was developed to cluster trajectories based on the distribution of entry points and trajectory vectors. Optimal number of trajectory clusters was determined by evaluating the fitted model via the elbow method as well as by displaying all clusters in the normalized space for visual inspection. In the dataset, the three most common histologic diagnoses were glioblastoma, brain metastases, and radiation necrosis. Stereotactic procedures performed included laser interstitial thermal therapy (LITT, n = 7 (5.79%)) and biopsy + LITT (n = 100 (82.6%)). There were no significant differences in the distribution of these variables between the discovery and validation cohorts. Unsupervised clustering revealed five canonical stereotactic trajectories: orbital-frontal, Kocher-frontal, mesial temporal, target-orthogonal, and cerebellar. Given the physical limitation of the MRI bore, each approach is associated with unique technical considerations in patient positioning, trajectory planning, and procedural execution. Across all approaches, the average radial error was consistently < 2 mm, with no statistically significant differences between canonical trajectories. We identified five reproducible, anatomy‑based canonical trajectories for ClearPoint LITT procedures performed within the MRI bore with the ClearPoint Prism system. These findings provide a foundation for standardizing stereotactic practice, improving workflow predictability, and supporting broader adoption MRI-guided stereotaxy across neuro‑oncologic centers.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

SURG-38 Canonical stereotactic trajectories and technical considerations for image-guided LITT — Science Skim