Case of implant failure following osteosynthesis for an intertrochanteric femoral fracture using a carbon fiber intramedullary nail.
Taisei Suzuki, Shigeo Fukunishi, Shohei Okahisa, Saori Niitsu, Syuya Matsumoto, Toshiya Tachibana
Department of Orthopedic Surgery, Takarazuka City Hospital, 4-5-1 Kohama-cho, Takarazuka City 665-0827, Hyogo, Japan. · Department of Orthopedic Surgery, Nishinomiya Kaisei Hospital, 1-4, Ohama-cho, Nishinomiya City 662-0957, Hyogo, Japan.
Key points
From the paper's conclusion- Ultimately, within just 6 months after the initial surgery, varus deformity developed at the fracture site, resulting in failure of the junction between the intramedullary nail and the lag screw, and necessitating revision surgery involving total hip arthroplasty.
Abstract
Carbon fiber-reinforced polyetheretherketone (CFR-PEEK) materials are radiolucent and produce fewer metallic artifacts during radiological examinations, making them a subject of growing interest as osteosynthesis materials in the field of orthopedics. However, clinical reports on carbon intramedullary nails are still limited, and there are few reports on postoperative complications. The present case report describes a 55-year-old female patient who sustained an unstable intertrochanteric fracture of the left femur and underwent open reduction and internal fixation using a CFR-PEEK intramedullary nail. Ultimately, within just 6 months after the initial surgery, varus deformity developed at the fracture site, resulting in failure of the junction between the intramedullary nail and the lag screw, and necessitating revision surgery involving total hip arthroplasty.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical evaluation of carbon fiber-reinforced polyether ether ketone versus titanium plates for fixation of high tibial osteotomy.Clinical biomechanics (Bristol, Avon) · 2026
- Utilization of Carbon Fiber Spinal Instrumentation in Primary Spinal Tumors: Outcome Analysis from the AO Spine Primary Tumor Research and Outcomes Network.Global spine journal · 2026
- Robot-Assisted Placement of Thoracic Carbon-Fiber-Reinforced Polyetheretherketone (CFR-Peek) Pedicle Screws in the Cervical Spine for Giant Cell Tumor: Technical NoteBioengineering · 2026
- Occult cement plug causing femoral canal obstruction during nailing after TKA: a case report.Frontiers in surgery · 2026
- Experimental and Numerical Investigation of Orthotropic CFR-PEEK and Ti-6Al-4V Osteosynthesis Plates Using a Multiscale Finite Element Modeling Approach.Polymers · 2026
- Cement Augmentation of the Blade in Proximal Femoral Nailing for Trochanteric Fractures in Elderly Patients: A Retrospective Comparison of Mechanical Stability and ComplicationsJournal of Clinical Medicine · 2025