Journal of nephrology

Feedback-controlled ultrafiltration as a useful tool for the diagnosis of volume status assessment in dialysis patients: a pilot study.

Veronica Cucci, Veronica Fortunato, Salvuccio Siciliano, Gabriele Carmelo Morreale, Pietro Suriano, Maria Eva Sberna, Maria Rita Stancanelli, Giuseppe Restivo,

Department of Medicine and Surgery, University of Enna "Kore", 94100, Enna, Italy. · Ward of Nephrology and Dialysis, Hospital "Umberto I", 94100, Enna, Italy.

Key points

From the paper's conclusion
  • Although further larger studies are needed, our pilot findings indicate potential exploratory utility for feedback-controlled ultrafiltration in assessing volume status among dialysis patients.

Abstract

BACKGROUND: Volume overload is one of the most critical challenges in dialysis patients and is monitored using bioelectrical impedance analysis (BIA). The feedback-controlled ultrafiltration (fUF) is used to reduce hypovolemic complications, but its diagnostic power has not been tested.Our analysis aims to evaluate the ability of the fUF to detect volume status alterations. METHODS: This study enrolled 29 dialysis patients. Volume status was defined by BIA as hypo/normo/hyper-volemia. The volume status assessment using the fUF was defined through the difference between the prescribed ultrafiltration and the effective ultrafiltration. Sensitivity, Specificity, Positive Predictive Value (PPV), Negative Predictive Value (NPV), and Accuracy were evaluated. The analysis was performed in subsamples based on the extracellular/intracellular (E/I) ratio of fluid distribution. RESULTS: For hypervolemia, the model showed a sensitivity of 83.3%, a specificity of 91.3%, a PPV of 71.4%, and an NPV of 95.5%. For hypovolemia the analysis showed a sensitivity of 70.0%, a specificity of 73.7%, a PPV of 58.3%, and an NPV of 82.4%. The correlation coefficient between BIA and fUF was 0.502 (p=0.006). Stratification by the E/I ratio revealed improved diagnostic performance in the subgroup with an E/I ratio > 0.8 (n=17). Global accuracy was 70.6%. The metrics for hypervolemia reached a sensitivity of 100% and an NPV of 100%, a specificity of 92.9%, a PPV of 75%, and a balanced Accuracy of 96.4%. CONCLUSIONS: Although further larger studies are needed, our pilot findings indicate potential exploratory utility for feedback-controlled ultrafiltration in assessing volume status among dialysis patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Feedback-controlled ultrafiltration as a useful tool for the diagnosis of volume status assessment in dialysis patients: a pilot study — Science Skim