Simplified weak-beam dark-field STEM method for bend-contour-free dislocation loop observation using a commercial STEM system.
Yusuke Noshi, Hiroshi Maeno, Moe Ishiwaki, Manato Kakizawa, Ryoya Ishigami, Kazuhiro Yasuda, Ken-Ichi Fukumoto
Graduate School of Engineering, University of Fukui, 1-3-33 Kanawa-cho, Tsuruga-shi, Fukui 914-0055, Japan. · The Ultramicroscopy Research Center, Kyushu University, 744 Motooka, Nishi-ku, Fukuoka 819-0395, Japan.
Key points
From the paper's conclusion- These improvements provide a more uniform background, enabling the reliable identification of the Burgers vectors of nanometer-scale dislocation loops (b = a < 100> and b = a/2<111>) through direct correlation of contrast behaviors across distinct g vectors.
Abstract
To advance the quantitative analysis of irradiation-induced defects, we successfully applied a simplified weak-beam dark-field scanning transmission electron microscopy (WBDF-STEM) technique using a standard spherical-aberration-corrected STEM setup, requiring no specialized hardware modifications. By systematically optimizing the illumination geometry with a 10 μm condenser aperture and the imaging geometry with camera lengths of 100 cm and 40 cm, we effectively suppressed macro-scale bend contours and micro-scale surface oxide moiré artifacts. This optimization relies on the combined effect of the probe convergence semi-angle and the acceptance semi-angle to integrate and average out localized intensity fluctuations, while objective-aperture selection at the 2g position reduces oxide-derived moiré contrast. These improvements provide a more uniform background, enabling the reliable identification of the Burgers vectors of nanometer-scale dislocation loops (b = a < 100> and b = a/2<111>) through direct correlation of contrast behaviors across distinct g vectors. Consequently, this methodology provides a practical and readily implementable approach for diffraction-contrast-based analysis of radiation-induced microstructures using a commercial AC-STEM instrument.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying the Full Damage Profile of Focused Ion Beams via 4D‐STEM Precession Electron Diffraction and PSNR MetricsSmall Methods · 2026
- Achieving 5-angstrom-resolution diffraction contrast with an uncorrected electron probe and nanosized defect characterizationsarXiv · 2026
- From phase analysis to phase tracking: a robust approach to atomic-scale strain mappingarXiv · 2026
- NLSTEM: Nonlocal Denoising for Enhanced 4D-STEM Pattern Indexing.Microscopy and microanalysis : the official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada · 2026
- Effective Point Spread Function Estimation for ADF-STEM Resolution Enhancement Using an Atomically Thin 2D Reference.Microscopy and microanalysis : the official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada · 2026
- Correlative 3D Electron Tomography and Nanobeam 4D-STEM Mapping of Orientation and Strain in Welded Silver Nanowires.Small (Weinheim an der Bergstrasse, Germany) · 2026