Evaluating Post-Analytical Stability of Blood Ethanol Samples Using Headspace-GC-FID.
David Cook, Jarrad Wagner
Nassau County Medical Examiner's Office - Department of Forensic Toxicology, 2251 Hempstead Turnpike, East Meadow NY, 11554. · Oklahoma State University CHS, 1111 W 17th St, Tulsa, OK 74107.
Key points
From the paper's conclusion- Over 6 months and 22 individual batch analyses, 303 samples were prepared and analyzed using Headspace-GC-FID for the presence of blood ethanol using 0.02% n-propanol as an internal standard.
Abstract
The analysis and determination of blood alcohol concentration (BAC) is a cornerstone test in forensic toxicology that is well researched. The purpose of this study was to determine the post analytical stability of blood ethanol obtained from re-injected vials via Agilent dual column headspace-GC-FID. Over 6 months and 22 individual batch analyses, 303 samples were prepared and analyzed using Headspace-GC-FID for the presence of blood ethanol using 0.02% n-propanol as an internal standard. Each individual batch was re-injected after storage at room temperature for 72 hours to mimic a typical weekend absence. Ethanol signal response, n-propanol signal response and measured blood ethanol concentration were evaluated for stability in terms of percent change. Sample population consisted of negative human blood fortified with known concentrations of ethanol via certified reference material purchased from accredited vendors (n = 187) in the form of calibrators and controls, as well as unknown post-mortem and ante-mortem blood samples submitted to the Nassau County Medical Examiner's Office Department of Forensic Toxicology (n = 116). All samples were tested using similar conditions and internal standard. Significant differences were observed in the signal response for both n-propanol (p<.001) and ethanol (p<.001) between the two test groups however, there was a non-significant difference in measured BAC between the two groups t(302) = 0.3, p = 0.780, d = 0.016. Results suggest very little effect of re-injection after 72-hour storage at room temperature on measured blood ethanol concentration using headspace GC-FID. While there was significant loss of ethanol and internal standard on reinjection, the calculated ethanol concentration was unaffected due to the similar properties of both volatile measurands as well as the consistency of the response ratio of ethanol:n-propanol in reinjected vials. In addition, the observed differences in measured BAC on re-injection were well within the NCME's measurement uncertainty of 11% at k = 3.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Forensic assessment of ethanol toxicity and validation of an analytical method for ethanol determination in blood and vitreous humor after autopsy.Legal medicine (Tokyo, Japan) · 2026
- Can Post-Mortem Cartilage Reliably Preserve Ethanol? A Time-Course Comparison with Blood and Urine.Toxics · 2026
- Determination of Alcohol in Blood by Headspace Gas Chromatography (HS-GC): A Systematic Review and Meta-Analysis.Journal of visualized experiments : JoVE · 2026
- Systematic Study on Phosphatidylethanol (PEth) Stability and De Novo Formation in Authentic Whole Blood Samples.Drug testing and analysis · 2026
- B-028 Review of a sample exchange program for postmortem vitreous biochemical analysisClinical Chemistry · 2026
- Extended blood ethyl glucuronide persistence after controlled ethanol administration: Implications for ethanol-negative delayed DUI cases.Forensic science international · 2026