Clinical ChemistryOpen accessReference values

Establishing Reference Values in Healthy Participants for the Cardiac Troponin T High-Sensitivity Gen 6 Assay: REF-TSIX Global Reference Study

Lori B Daniels, Christian Mueller, Evangelos Giannitsis, Steven J. R. Meex, David Buehlmann, Peter Dilba, Garnet Bendig, Mette Cole,

University of California San Diego · University Hospital of Basel

Observational · n=8,294

Key points

From the paper's conclusion
  • This study determined sex-specific and uniform 99th percentile URLs for the Troponin T high-sensitivity Gen 6 assay that were comparable irrespective of the matrix used in a large, global, healthy reference population.

Abstract

BACKGROUND: Measurement of cardiac troponin (cTn) using high-sensitivity assays is recommended for the diagnosis of myocardial infarction. We determined sex-specific and uniform 99th percentile upper reference limits (URLs) using the new Elecsys® Troponin T high-sensitivity Gen 6 assay in a global, healthy reference range cohort. METHODS: Lithium-heparin plasma and serum samples were prospectively collected from apparently healthy individuals aged ≥20 years across 34 global sites in the United States, Europe, China, and Japan. cTnT was measured using the Troponin T high-sensitivity Gen 6 assay on the Cobas® e 801 analyzer. Exclusion criteria were defined according to the 2022 IFCC guidance. Uniform and sex-specific 99th percentile URLs and nonparametric 95% CIs were determined in plasma and serum separately and combined. RESULTS: The final study population comprised 4147 participants (52.5% female) with a median (25th-75th percentiles) age of 48.0 (33.0-59.0) years; 45.8%, 47.9%, 5.2%, and 1.1% were White, Asian, Black, and other/unknown, respectively. For sample matrices combined (n = 8294), 81.0% and 99.2% of cTnT values were above the limit of detection in females and males, respectively. Sex-specific 99th percentile URLs (95% CI) were 18 (16-23) ng/L for females and 32 (28-35) ng/L for males; the uniform 99th percentile URL was 27 (24-31) ng/L. URLs were comparable in plasma and serum samples. CONCLUSIONS: This study determined sex-specific and uniform 99th percentile URLs for the Troponin T high-sensitivity Gen 6 assay that were comparable irrespective of the matrix used in a large, global, healthy reference population.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Establishing Reference Values in Healthy Participants for the Cardiac Troponin T High-Sensitivity Gen 6 Assay: REF-TSIX Global Reference Study — Science Skim