Performance of an infliximab precision-guided dosing tool from a capillary blood-based self-collection device in inflammatory bowel disease patients.
Cindy C Y Law, Arafa Djalal, Mason Hopkins, Dakota Trick, Thierry Dervieux, Catherine Rowan, Jean-Frederic Colombel
Division of Gastroenterology, Department of Medicine, Massachusetts General Hospital, Boston, MA, United States. · Susan and Leonard Feinstein Inflammatory Bowel Disease Clinical Center, Division of Gastroenterology, Mount Sinai Hospital, New York, NY, United States.
Key points
From the paper's conclusion- PredictrPK accurately predicted IFX trough concentrations. Capillary self-collection using an upper-arm device yielded results comparable to those from venous blood measurement, which may facilitate increased implementation and accessibility of PGD in clinical practice.
Abstract
BACKGROUND AND AIMS: Infliximab (IFX) trough concentrations are routinely measured to guide dosing and optimize efficacy. Bayesian forecasting tools may enable earlier prediction of trough concentrations, while patient-friendly capillary self-collection devices could improve access to precision-guided dosing (PGD). We evaluated the performance of the PredictrPK IFX platform in forecasting IFX trough concentrations by using serum collected with an upper-arm self-collection device and compared its performance with standard venous blood obtained by phlebotomy. METHODS: Patients with inflammatory bowel disease (IBD) receiving IFX maintenance therapy were recruited from a single IBD center. Midcycle samples collected via an upper-arm self-collection device and via traditional venipuncture were compared. The IFX clearance was estimated by use of a nonlinear mixed-effects model with Bayesian priors and the predicted IFX levels were compared with measured trough concentrations obtained prior to infusions. Deming regression was used for statistical analysis. RESULTS: The study included 57 patients (61% males, median age, 34 years). The median IFX dose was 10 mg/kg with a median interdose interval of 6 weeks. A total of 97 capillary and venous blood pairs were tested. Concentrations of IFX from venous and capillary blood samples demonstrated strong correlations (R2 = 0.987, Demings = 1.08). There was also a strong association between predicted IFX concentrations and measured IFX trough concentrations. CONCLUSION: PredictrPK accurately predicted IFX trough concentrations. Capillary self-collection using an upper-arm device yielded results comparable to those from venous blood measurement, which may facilitate increased implementation and accessibility of PGD in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Baseline Clearance and Early Troughs Identify Candidates for Early Therapeutic Drug Monitoring During Infliximab Induction in Inflammatory Bowel Disease.Alimentary pharmacology & therapeutics · 2026
- Population Pharmacokinetics of Subcutaneous Infliximab: A Real-World Study of Naïve and Switch Patients.Pharmaceutics · 2026
- Practical antimicrobial TDM (PAT) of vancomycin in Japanese emergency and critical care units: Concentration overestimation and correction using scaling factors.International journal of clinical pharmacology and therapeutics · 2026
- Connecting the Dots Between Pharmacokinetics, Inflammatory Biomarkers, and Mucosal Healing in Patients on Infliximab with Inflammatory Bowel Disease.Pharmaceutics · 2026
- Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.Journal of separation science · 2026
- Perianal fistulizing Crohn's disease increases infliximab clearance and causes low infliximab exposure (ATLANTIC).Journal of Crohn's & colitis · 2026