Distinct intracardiac electrogram waveforms with perforation during left bundle branch area pacing implantation
Heli Tolppanen, Valerian Valiton, Samuel Stempfel, Haran Burri
Helsinki University Hospital · University Hospital of Geneva
Key points
From the paper's conclusion- Perforation during left bundle branch area pacing (LBBAP) results in a fall in the current of injury (COI) amplitude in the unipolar unfiltered electrogram (iEGM), but systematic waveform analyses have not been performed.
Abstract
AIMS: Perforation during left bundle branch area pacing (LBBAP) results in a fall in the current of injury (COI) amplitude in the unipolar unfiltered electrogram (iEGM), but systematic waveform analyses have not been performed. Our aim was to investigate unipolar iEGM waveforms during perforation and to compare them to those at the final lead position. METHODS AND RESULTS: The iEGMS of consecutive patients who had perforation during LBBAP implantation were systematically analysed. A total of 92 patients with perforation were included. In the unfiltered channel, sensed COI amplitude was lower with perforation [3.0 (1.5-4.1) mV] than at the final lead position [14.0 (9.2-17.5) mV, P < 0.0001], as was also the case during pacing. Patients with narrow QRS/non-LBBB typically had wide negative (QS) waveforms during sensing (in 67% of cases), whereas those with LBBB/paced rhythm had positive (wide R/RS) morphologies (in 93% of cases). In the former subgroup, a sensed Q or S amplitude > COI amplitude (which is easy to eyeball during lead deployment) had a sensitivity of 86% and a specificity of 93% for diagnosing perforation. Waveforms during macroperforation (with loss of capture, n = 27) differed compared to microperforation (with preserved capture, n = 65), with significantly lower COI amplitudes, more frequent QS morphology, and rarer sharp multiphasic components in the ventriculogram of the filtered channel. CONCLUSION: Beyond COI amplitude, additional iEGM waveform parameters may be used to evaluate the presence of LBBAP perforation and should be carefully monitored during lead deployment to improve safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Usefulness of the Current of Injury Function in a Pacing System Analyzers for Left Bundle Branch Area Pacing Implantation : A Case Report.Indian pacing and electrophysiology journal · 2026
- Effects of LBBAP Pacing-Site Position under a Modelled Proximal LBBB: A 24-Model Electrophysiological Simulation Study.Medical engineering & physics · 2026
- Evaluation of a novel pacemaker-based electrogram morphology monitor algorithm in conduction system pacingEuropean Heart Journal · 2026
- Left Bundle Branch Capture During Left Bundle Branch Area Pacing: A Diagnostic Accuracy Systematic Review.Pacing and clinical electrophysiology : PACE · 2026
- Optimized Strategy for Improving the Accuracy of Determining Left Bundle Branch Capture.Pacing and clinical electrophysiology : PACE · 2026
- Stepwise Electrosurgical Bailout During Difficult Left Bundle Branch Area Pacing.Pacing and clinical electrophysiology : PACE · 2026