Where do leadless pacemakers really land? Insights from a CT validated multicenter investigation
K Abdelsayed, J L Cavalcante, A Bahbah, M T Desouki, D Witt, E Halvorson, M Katsiyiannis, M J Kapphahn-Bergs,
Key points
From the paper's conclusion- Compared with septal placement, non-septal implantation was associated with lower female prevalence (15% vs 43%; p<0.001), greater RV height (78±13.6 vs 70±12.2 mm; p<0.01), and increased distance from the tricuspid annulus to device tip (33±11.8 vs 25±12.3 mm; p<0.01) (Table 1).
Abstract
Background Optimal safety and performance of leadless pacemakers (LPMs) depend on mid-to-apical septal placement, whereas non-septal positions increase risks of perforation, dyssynchrony, and pacing-induced cardiomyopathy. Fluoroscopic localization is often unreliable with spatial localization. The accuracy and implications of device deployment with the current standards of practice is still poorly identified. Objective We evaluated the accuracy of procedure-note–reported LPM location against CT-verified position and assessed the association of the CT-adjudicated positions with baseline characteristics, right ventricular (RV) measurements, and paced QRS duration. Methods We included consecutive patients who underwent LPM implantation and contrast-enhanced CT of optimal quality at three tertiary centers (2015–2024). The right ventricle was segmented into mid/high-septal, apical-septal, apical, apical free-wall, and mid/high free-wall regions. CT-verified location (two independent readers) was compared with procedure-note–reported position. Baseline characteristics, CT measurements, and paced QRS were compared between septal and non-septal locations. Results Of 540 screened LPM patients, 92 were eligible (mean age 76±9.8 years, 31% females; Table 1). Mean paced QRS duration was 174 ms (±16.9), with 72% having a QRS duration >160 ms. About 42.5% had ≥1-segment discordance between procedure note and CT location (Figure 1). Twenty-seven percent had the device implanted on the apical (17%) or mid/high (10%) free-wall (Figure 2). CT-verified non-septal pacing showed significantly wider paced QRS than septal pacing (180±20 vs 166±12 ms; p<0.0001). Compared with septal placement, non-septal implantation was associated with lower female prevalence (15% vs 43%; p<0.001), greater RV height (78±13.6 vs 70±12.2 mm; p<0.01), and increased distance from the tricuspid annulus to device tip (33±11.8 vs 25±12.3 mm; p<0.01) (Table 1). Conclusion Nearly half of LPM implants exhibited misclassification of device position when compared with CT. Non-septal and free wall placement were common and showed a substantially wider paced QRS, supporting the need for improved intraprocedural spatial guidance. Selective integration of advanced imaging may enhance localization accuracy and optimize LPM outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Septal-Inferior Vena Cava Angle and Accurate Septal Implantation of Leadless Pacemakers: Computed Tomography-Based Study.Pacing and clinical electrophysiology : PACE · 2026
- Relationship between electrical dyssynchrony and left ventricular function in patients with leadless pacemaker; insights from myocardial work analysisEuropean Heart Journal · 2026
- Preprocedural Computed Tomography Guidance Improves Accuracy of Right Ventricular Septal Lead Placement in Pacemaker Implantation.Journal of cardiovascular electrophysiology · 2026
- Leadless Pacemaker Implant Location as a Predictor of Pacing Induced Cardiomyopathy.Pacing and clinical electrophysiology : PACE · 2026
- Comparative electrophysiology of aortic root pacing in humansEuropean Heart Journal · 2026
- High right ventricular pacing burden is associated with heart failure following leadless pacemaker implantationsEuropean Heart Journal · 2026