Autonomous closed-loop intravenous esmolol for systolic blood pressure regulation and beat-to-beat variability suppression with safety supervision: a large-animal validation study
J Curneen, J O'Dwyer, C Judge, D Graham, S A Cryan, E Roche, G Duffy
Key points
From the paper's conclusion- Conclusions A closed-loop intravenous esmolol system achieved autonomous SBP regulation within a predefined target band while simultaneously reducing beat-to-beat BPV, with preserved secondary haemodynamic stability and reliable automated safety responses in a large-animal model.
Abstract
Background/Introduction Acute hypertension management in high-acuity settings - including hypertensive emergencies and perioperative care - relies on manual drug titration, which is subject to operator delays, dose imprecision, and haemodynamic overshoot.1 Beyond threshold attainment, blood pressure variability (BPV) is increasingly recognised as an independent cardiovascular risk factor, with the contribution of short-term beat-to-beat fluctuations to haemodynamic stress in the acute setting requiring further exploration.2,3,4 Autonomous closed-loop control offers the potential to achieve both precise target attainment and BPV minimisation simultaneously, but preclinical validation with integrated safety supervision is lacking. Purpose To develop and validate a closed-loop intravenous esmolol infusion system for automated systolic blood pressure (SBP) regulation and BPV reduction in a large-animal model, incorporating real-time safety supervision and pharmacological rescue. Methods Following institutional ethical approval, five female Yorkshire swine underwent intravenous esmolol infusion across five experimental days. Open-loop dose-response characterisation (n=1) informed controller design. Closed-loop control employed a multi-layer feedback architecture with adaptive modulation to accommodate evolving haemodynamic conditions and variability dynamics. An independent supervisory safety layer provided autonomous override and pharmacological counter-regulation in the event of excessive blood pressure reduction. A predefined SBP reduction band was specified. Performance metrics included time-in-range (TIR) and average real variability (ARV). Results In closed-loop validation experiments, representative performance demonstrated TIR of 51.7% versus 0% during the untreated phase. During steady-state regulation, ARV was reduced by 31% (0.91 vs 1.32 mmHg), confirming simultaneous target attainment and beat-to-beat variability suppression. Diastolic blood pressure and heart rate decreased by 6.0% and 9.1% respectively, remaining within safe limits. In dedicated safety stress-testing, four autonomous SBP threshold violations triggered immediate pump suspension and drug rescue with successful controller re-engagement, without operator input. Conclusions A closed-loop intravenous esmolol system achieved autonomous SBP regulation within a predefined target band while simultaneously reducing beat-to-beat BPV, with preserved secondary haemodynamic stability and reliable automated safety responses in a large-animal model. These findings provide proof-of-concept for device-based autonomous hypertension and BPV control warranting translation to human studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Automatic control of arterial blood pressure: computer, benchtop, and animal models.Frontiers in medical technology · 2026
- Clinical Use of an Automated SBP-Responsive Pneumatic Tourniquet: A Single-Center Retrospective Study.Journal of visualized experiments : JoVE · 2026
- Early blood pressure variability and successful extubation in critically ill patients: a prospective observational studyEuropean Journal of Medical Research · 2026
- A risk-aware individualized blood pressure target to avoid hypotension in surgeryarXiv · 2026
- Evolutionary Optimization of Extended Kalman Filter Hyperparameters for Adaptive Data-Driven ABP Prediction Framework Inspired by Cardiovascular Digital-Twin PrinciplesAlgorithms · 2026
- Automated epidural spinal cord stimulation for cardiovascular regulation in spinal cord injury: from optimization to real-time implementation.medRxiv : the preprint server for health sciences · 2026