Towards understanding the discordance between invasive and echocardiographic pressure gradients after transcatheter aortic valve implantation
K Hovde, J M Aalen, C H Eek, A Opdahl, S S Clementsen, E W Remme, P S Halvorsen, J P E Kvitting,
Key points
From the paper's conclusion- Conclusion Echocardiography systematically overestimates transvalvular mean pressure gradients compared with invasive measurements after TAVI, but simultaneous assessment substantially reduced the discordance.
Abstract
Background Echocardiography at discharge frequently estimates higher transvalvular mean pressure gradients (MG) compared to intraoperative invasive measurements following transcatheter aortic valve implantation (TAVI). This discordance is incompletely understood and may lead to misinterpretation of valve performance and unwarranted invasive procedures and re-interventions. Purpose To assess the extent of discordance between invasive intraoperative pressure gradients and echocardiography-derived pressure gradients. Furthermore, to test the hypothesis that procedural interventions such as conscious sedation and rapid pacing may reduce transvalvular flow during the TAVI-procedure and hence explain the discordance. Methods In this prospective study, 34 patients underwent simultaneous invasive and transthoracic echocardiographic pressure gradient measurements. Assessment of simultaneous gradients was performed immediately after TAVI, both at baseline and during pharmacologically increased flow using dobutamine. Echocardiography was repeated the following day (discharge). Results Echocardiographic MG at discharge was significantly higher than invasive MG measured immediately post-TAVI (12 ± 4 mmHg vs. 4 ± 3 mmHg, p < 0.001; mean difference 8 ± 5 mmHg). In contrast, simultaneous measurements immediately post-TAVI reduced, but did not eliminate the discordance (mean difference 3 ± 4 mmHg, p < 0.001). Dobutamine infusion increased both echocardiographic and invasive MG (8 ± 5 to 13 ± 4 mmHg and 4 ± 3 to 8 ± 5 mmHg, respectively, both p < 0.001). Echocardiographic MG was similar during dobutamine infusion and at discharge (13 ± 4 vs. 12 ± 4, p = 0.24). A representative patient example is shown in the Figure. Conclusion Echocardiography systematically overestimates transvalvular mean pressure gradients compared with invasive measurements after TAVI, but simultaneous assessment substantially reduced the discordance. This may be explained by differences in transvalvular flow.FigureFor image description, please refer to the figure legend and surrounding text.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transmitral mean pressure gradient after transcatheter aortic valve replacement and late mortalityEuropean Heart Journal · 2026
- Invasive and echocardiographic gradients of self- and balloon-expandable valves in failing aortic bioprostheses.Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation · 2026
- Prognostic value of early haemodynamic valve deterioration after TAVIEuroIntervention · 2026
- Different Types of Aortic Valve Stenosis in Patients Undergoing Transcatheter Aortic Valve ReplacementStructural Heart · 2026
- Early Post-transcatheter Aortic Valve Implantation (TAVI) Haemodynamics With SAPIEN Balloon-Expandable and Evolut Self-Expanding Valve Platforms: A Retrospective Single-Centre Cohort Study.Cureus · 2026
- Real-Time Hemodynamic Guidance During Valve-in-Valve TAVI.JACC. Case reports · 2026