Post-implant CT volumetry shows frequent divergence between segmented and nominal factory valve volume after TAVI
D Hochstein, E Ben-Assa, R Nahmias-Oz, N Nagar, O Vazhgovsky, Y Steinmetz, S Tejman-Yarden, D Dvid,
Key points
From the paper's conclusion- Conclusions Post-implant CT demonstrates frequent and quantitatively substantial divergence between nominal factory valve volume and in vivo segmented valve volume after TAVI, most often in the direction of expansion.
Abstract
Background Transcatheter aortic valve devices are manufactured to a nominal geometry, yet deployment within a patient-specific annulus and aortic root can alter frame morphology. However, systematically acquired post-implant cardiac CT datasets after TAVI are uncommon, and consequently only limited published data exist that quantitatively characterize real-world, in vivo valve geometry after implantation. Leveraging a structured post-implant CT cohort, we quantified the difference between manufacturer-derived valve volume and in vivo post-implant valve volume using STL-based volumetric analysis from CT reconstructions. Purpose To compare post-implant CT-derived segmented valve volume with nominal factory valve volume and quantify the magnitude, direction, and variability of volumetric deviation in vivo. Methods We retrospectively analyzed 75 patients who underwent post-implant cardiac CT. Three-dimensional valve models were reconstructed from CT DICOM datasets using dedicated cardiovascular imaging software, exported as STL meshes, and valve volumes were computed in mm³. These in vivo volumes were compared with the corresponding manufacturer (factory) valve volumes for each valve size. We evaluated signed and absolute volume differences and summarized relative (%) change. Paired testing assessed whether mean signed differences differed from zero. Results Among 75 patients, segmented valve volume exceeded factory volume in 61 (81.3%), was lower in 13 (17.3%), and was identical in 1 (1.3%). Mean segmented valve volume was 2060.0 mm³ versus 1649.6 mm³ for factory volume. Mean signed difference was +410.4 mm³ (95% CI 311.4–509.3; p<0.001). Mean absolute difference was 470.5 mm³, with a median of 401 mm³ and a range of 0–1629 mm³. Mean relative change was +25.9% (median +23.1%). Absolute deviation was substantial across all nominal valve sizes, while proportional change was greatest in smaller valves, with mean relative increases of 35.1% for ~1243 mm³ valves, 25.7% for ~1569 mm³ valves, and 20.2% for ~1977 mm³ valves. Conclusions Post-implant CT demonstrates frequent and quantitatively substantial divergence between nominal factory valve volume and in vivo segmented valve volume after TAVI, most often in the direction of expansion. Smaller nominal valves show the greatest proportional change. CT-based volumetry may provide a practical framework for understanding post-deployment valve geometry in real-world patientsSegmented ValvesSegmented Valves on Factory Valves
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-Time Fluoroscopic Assessment of Underexpansion Predicts Hemodynamic Valve Deterioration Following TAVR With Balloon-Expandable DeviceJACC: Cardiovascular Interventions · 2025
- Valve Embolisation During Transcatheter Aortic Valve Implantation (TAVI): Clinical Outcomes and CT-Derived Anatomical Characteristics.Cureus · 2026
- In Silico Modeling and Validation of Post-Dilation in TAVI Patients.Annals of biomedical engineering · 2026
- Comparative analysis of TAVI planning software: implications for prosthesis sizing and clinical outcomeEuropean Heart Journal · 2026
- In silico modeling of transcatheter heart valve oversizing and ellipticity, Part I: Establishing credibility of an advanced modelComputer Methods and Programs in Biomedicine · 2026
- Cardiac magnetic resonance evaluation of valve function after Sapien versus Myval transcatheter aortic valve implantation: A COMPARE-TAVI 1 substudyEuropean Heart Journal - Valvular and Structural Heart Disease · 2026