European Heart Journal

Unlocking 360° multiplane transoesophageal echocardiography for transseptal puncture guidance in mitral transcatheter edge-to-edge repair

M Stapor, K Golinska Grzybala, B Szlosarczyk, J Trebacz, K Szostek, P Piatek, T Dziwinski, A Piorkowski,

Observational · n=211

Key points

From the paper's conclusion
  • Enabling optional unrestricted 360° multiplane rotation may facilitate more efficient transseptal guidance during structural heart interventions.

Abstract

Background Contemporary matrix-array transoesophageal echocardiography (TOE) probes technically allow full 360° multiplane rotation; however, clinical imaging remains restricted to the conventional 0–180° range. This limitation may impair procedural efficiency during transseptal puncture (TSP) in mitral transcatheter edge-to-edge repair (M-TEER). Purpose To assess the frequency of mirrored four-chamber (4Ch) imaging (>140°) during TSP in M-TEER and its association with procedural timing. A secondary objective was to evaluate the feasibility of unrestricted 360° multiplane imaging in a simulation environment. Methods We performed a retrospective analysis of 211 consecutive patients undergoing M-TEER. TOE recordings were reviewed to identify cases requiring mirrored 4Ch imaging during TSP. TSP time was defined as the interval from right atrial instrumentation to septal puncture. Additionally, a TOE simulator was modified to allow unrestricted 360° multiplane rotation, enabling negative-angle imaging (<0°). Results Mirrored 4Ch imaging was required in 30 cases (14.2%). Median TSP time was 394 [291–560] seconds. Procedures requiring mirrored imaging were associated with significantly longer TSP duration compared with standard imaging (502 [326–745] vs. 383 [280–514] seconds; p = 0.039). Unrestricted 360° multiplane rotation was technically feasible in a simulator, enabling acquisition of mirrored and negative-angle views. Conclusion Mirrored 4Ch imaging during TSP is a frequent and clinically relevant phenomenon associated with prolonged transseptal access in M-TEER. Enabling optional unrestricted 360° multiplane rotation may facilitate more efficient transseptal guidance during structural heart interventions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers