European heart journal. Valvular and structural heart disease

Latest-iteration self-expanding vs balloon-expandable transcatheter heart valves for failed surgical bioprostheses: The INNOVATE registry.

Antonio Sisinni, Antonio Popolo Rubbio, Azeem Latib, Ignacio J Amat Santos, Christina Brinkmann, Cristina Aurigemma, Claudio Montalto, Damiano Regazzoli,

Department of Cardiology, IRCCS Policlinico San Donato, Piazza Edmondo Malan 2, San Donato Milanese, Milan 20097, Italy. · Montefiore-Einstein Center for Heart and Vascular Care Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, NY, USA.

Observational · n=381

Key points

From the paper's conclusion
  • Latest iterations of SE and BE THVs provided favourable outcomes in ViV TAVI, without significant differences in safety/efficacy profile. SE platforms achieved lower residual gradients at 1 year.

Abstract

BACKGROUND: Valve-in-valve (ViV) transcatheter aortic valve implantation (TAVI) is increasingly performed in degenerated surgical bioprostheses, although data concerning the performance of latest iterations of transcatheter heart valves (THVs) are scarce. AIMS: We aimed to compare procedural, clinical and haemodynamic outcomes of latest-generation THVs in ViV TAVI. METHODS AND RESULTS: This multicenter retrospective study included 381 consecutive patients undergoing ViV TAVI, 294 with self-expanding (SE), and 87 with a balloon expandable (BE) THV. Primary efficacy endpoints were Valve Academic Research Consortium (VARC)-3 technical and device success. Primary safety outcomes were all-cause death and re-hospitalization for heart failure (HF). Patients receiving SE valves were older (79 vs 75 years, P < .001). Time from surgery to ViV was longer in BE group (4 vs 3 years, P = .041). A numerically higher rate of coronary obstruction (4% vs 0%, P = .076) was observed in SE THVs, despite a more frequent coronary protection (24% vs 14%, P = .045). Technical (95% vs 98%, P = .380) and device success (89% vs 79%, P = .105) were similar. At 1 year, SE valves showed significantly lower mean gradients (14 ± 8 vs 21 ± 8 mmHg, P < .001). Rates of paravalvular leak and patient-prosthesis mismatch did not differ between groups. All-cause mortality (5% vs 4%, P = .928) and re-hospitalization for HF (4% vs 2%, P = .507) was also comparable among SE and BE, likewise myocardial infarction, stroke, and permanent pacemaker implantation. CONCLUSION: Latest iterations of SE and BE THVs provided favourable outcomes in ViV TAVI, without significant differences in safety/efficacy profile. SE platforms achieved lower residual gradients at 1 year.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Latest-iteration self-expanding vs balloon-expandable transcatheter heart valves for failed surgical bioprostheses: The INNOVATE registry — Science Skim