Assessment of cardiac image quality from single-probe handheld ultrasound devices.
Anna Grimaldo, Benjamin Koethe, Shane Williams, Eileen W Mai, Joseph Nguyen, Monica M Dehn, Lisa Small, Jennie An,
Department of Medicine, Tufts Medical Center, Boston, MA, USA. · Institute for Clinical Research and Health Policy Studies, Tufts Medical Center, Boston, MA, USA.
Key points
From the paper's conclusion- Conclusions Basic interpretable imaging from single-probe handheld ultrasound devices can be obtained for most patients under idealized conditions, suggesting possible use for enhanced cardiac disease detection.
Abstract
AIMS: This study aims to assess quality of cardiac ultrasound imaging from commercially available single-probe handheld devices. METHODS AND RESULTS: Cardiac imaging from single-probe handheld ultrasound devices was prospectively acquired by registered cardiac sonographers. Videos were graded on a scale of 1 (uninterpretable) to 5 (diagnostic-quality imaging) by 2 sonographers. The average image quality score for each cine loop was analysed. Interpretable imaging was defined as an average IQ score ≥ 2. Good image quality was defined as an average IQ score ≥ 4. Imaging from single-probe devices representing 518 videos from 206 patients was assessed. The mean age was 72 (SD 11.1) years. Agreement on image quality was moderate (weighted kappa 0.50, 95% CI 0.45-0.55). The correlation between sonographer quality grades was 0.60. The median average IQ score was 2.5 (1.5). 454 (88%) of the videos were interpretable (average IQ ≥ 2). At least one interpretable video was obtained for 187 (91%) patients. 62/518 (12%) of videos had good image quality (average score ≥ 4). There was a trend towards lower image quality with increasing BMI categories [16-25, median 3.0 (2.0, 3.5), 25-29, median 3.0 (2.0, 3.5), > 29 median 2.5 (2.0, 3.0), P < 0.001]. CONCLUSION: Basic interpretable imaging from single-probe handheld ultrasound devices can be obtained for most patients under idealized conditions, suggesting possible use for enhanced cardiac disease detection. Since imaging from a significant minority of patients is uninterpretable and lower image quality is associated with higher BMI, rigorous assessment of image quality is essential prior to image interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia.Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2026
- Automated artificial intelligence-enabled measurement of cardiac structures on point-of-care ultrasonography: a prospective multicenter validation study.Frontiers in cardiovascular medicine · 2026
- The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System.Diagnostics (Basel, Switzerland) · 2026
- Novice users of handheld cardiac ultrasound: comparative performance of two artificial intelligence guidance systemsEuropean Heart Journal · 2026
- Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assuranceJournal of Applied Clinical Medical Physics · 2026
- AI-assisted handheld echocardiography for hospital bedside cardiac triage: The prospective OPTIMUST implementation study.International journal of cardiology · 2026