Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout.
Anh Do Nguyen, Tien Hoa Phan, Son Truong Do Chau
Department of Interventional Cardiology, Nhan dan Gia Dinh Hospital, 1 No Trang Long Street, Gia Dinh Ward, Ho Chi Minh City 700000, Vietnam.
Key points
From the paper's conclusion- A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the
Abstract
BACKGROUND: Balloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously. CASE SUMMARY: A 62-year-old man with multivessel coronary disease underwent staged PCI of a left circumflex-obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully. DISCUSSION: Sequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bailout Retrieval of a Ruptured Entrapped Coronary Balloon Using the Reinflation-Deflation Technique (Nantapon-Wittawat Technique).Cureus · 2026
- Reverse Balloon Anchoring for Retrieval of an Entrapped Side-Branch Balloon During the Jailed Balloon Technique.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Case report: successful retrieval of an entrapped balloon during balloon pericardiotomy after rupturing.European heart journal. Case reports · 2026
- Endovascular Retrieval of Unraveled Coils Extending Along the Aorta Due to Guidewire Fracture During Percutaneous Coronary Intervention: The Snare and Trap Technique.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Bailout rotational atherectomy rather than repeat balloon inflation after balloon rupture with extensive dissection and Ellis type II perforation in an undilatable calcified lesion.Annales de cardiologie et d'angeiologie · 2026
- Gutter Rotation Technique for Percutaneous Ipsilateral Retrieval of a Ruptured SAPIEN-3 Balloon.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026