B-414 Analytical Performance Evaluation of a Cyclosporine Assay Under Development for the Alinity i Platform
Himali Soni, Manu Sankar, Nicholas Gavigan, Jose Ortiz, Salman Ali, Susan Brophy
Key points
From the paper's conclusion- Conclusion The Cyclosporine assay under development for the Alinity i platform demonstrated robust analytical performance and strong concordance with the on-market ARCHITECT Cyclosporine assay, supporting its suitability as a reliable option for cyclosporine monitoring on the Alinity i system for the U.S. market.
Abstract
Background Cyclosporine is a fungal-derived cyclic undecapeptide and a potent immunosuppressant drug used widely in solid organ transplantation. It inhibits T-cell activation by blocking IL-2 transcription, improving survival in heart, liver, and kidney transplants. As its absorption is variable and bioavailability changes over time, therapeutic drug monitoring is essential. Cyclosporine is extensively metabolized by hepatic CYP-450 enzymes, and its levels are affected by numerous drug interactions. Despite its benefits, it carries significant risks, including nephrotoxicity, hepatotoxicity, and gastrointestinal and neurological side effects, making regular monitoring crucial to minimize complications such as acute rejection. Development of the Cyclosporine assay on the Alinity i platform will expand the system’s transplant drug monitoring menu by providing Cyclosporine testing alongside existing Sirolimus and Tacrolimus assays within the Alinity i portfolio for the U.S. market. Methods Cyclosporine levels in whole blood were measured using a two-step chemiluminescent microparticle immunoassay (CMIA) performed on the Alinity i system. Prior to automated Alinity i analysis, samples underwent manual pretreatment. Whole blood specimens were lysed with a solubilization reagent, mixed with a precipitation reagent for the extraction of cyclosporine, and centrifuged. The resulting supernatant was decanted into a transplant pretreatment tube and loaded onto the analyzer. In the assay sequence, the pretreated sample, assay diluent, and anti-cyclosporine–coated paramagnetic microparticles were combined, allowing cyclosporine to bind to the microparticles. After washing, an acridinium-labeled cyclosporine conjugate was added. Following a second wash and the addition of Pre-Trigger and Trigger solutions, a chemiluminescent reaction was generated. The Alinity i system measures relative light units (RLUs), which are inversely proportional to the cyclosporine concentration. Results The Alinity i Cyclosporine assay under development demonstrated analytical performance comparable to the Cyclosporine assay on the ARCHITECT platform. The assay has a limit of blank, limit of detection and limit of Quantitation of = 1.1ng/mL, = 5.6 ng/mL and = 12.0 ng/mL, respectively. Total within laboratory imprecision was observed to be = 9.7% for samples between 40 ng/mL to 1500 ng/mL and SD of = 6.0 ng/mL for samples < 40ng/mL. The assay demonstrated linearity up to 1725.7 ng/mL and has reagent onboard stability of at least 14 days. Method comparison to LC-MS showed a regression slope of 1.04 and correlation coefficient of 0.988. Comparison to the on-market ARCHITECT assay demonstrated a regression slope of 0.94 and correlation coefficient of 0.994 with < 10% predicted bias at the medical decision points 50, 100, and 800 ng/mL. The Cyclosporine assay under development for the Alinity i platform demonstrated accuracy by recovery between 100%±10% for samples across the analytical measuring interval. Conclusion The Cyclosporine assay under development for the Alinity i platform demonstrated robust analytical performance and strong concordance with the on-market ARCHITECT Cyclosporine assay, supporting its suitability as a reliable option for cyclosporine monitoring on the Alinity i system for the U.S. market.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-405 Estimation of reference change value for the cyclosporine assay and implementation of delta check auto verificationClinical Chemistry · 2026
- A high-throughput LC-MS/MS assay for therapeutic drug monitoring of combined immunosuppressive and antimicrobial therapy.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Cyclosporine A, a macrocyclic peptide, is transported by organic anion transporting polypeptide 1B1/1B3 at clinically relevant doses.Drug metabolism and disposition: the biological fate of chemicals · 2026
- Does CytoSorb Interfere with Immunosuppression? A Pharmacokinetic and Functional EvaluationPharmaceutics · 2025
- Cyclosporine A revisited: Molecular mechanisms, non-immune physiological pathways, and the architecture of multi-organ toxicities.Current opinion in pharmacology · 2026
- Correlation analysis of cyclosporine A trough concentration with hematological and biochemical parameters: a retrospective study in pediatric aplastic anemia.Frontiers in pediatrics · 2026