B-398 Efficiency and scalability unlocked: multi-channel liquid chromatography integrated with a high-resolution accurate mass time-of-flight mass spec as a new platform for LDTs

Paula Ladwig, Mindy Kohlhagen, Chris Thompson, Anthony Maus, David Murray, Maria Alice Willrich

Key points

From the paper's conclusion
  • We were successfully able to run across 2 channels using the 10-minute LC method with a 3-minute MS window to allow for multiplexing.

Abstract

Background Liquid chromatography mass spectrometry (LC-MS) has become a common platform for the implementation of lab-developed tests (LDTs). MS can offer a fast pathway to assay development with low cost-per-assay, multiple assays per platform, and no costly analyte specific reagents. Therapeutic monoclonal antibodies (tmAb) used in the treatment of autoimmune disorders and cancers are routinely measured by MS for therapeutic drug monitoring. Recently high-resolution accurate mass (HRAM) MS has been proven useful for the detection of tmAbs and also endogenous monoclonal antibodies in multiple myeloma minimal residual disease (MRD). While our tmAb and MRD development was performed using a microflow single-channel LC integrated with a Sciex TripleTOF 5600, we looked to instrumentation with higher capacity for implementation; triple quads or orbitraps that could be integrated with multiplexing, ultra-high-performance liquid chromatography (UHPLC) allowing for throughput and savings in turnaround time (TAT). Our goal was to investigate the possibility of integrating a multiplexing 2-channel UHPLC with a HRAM MS for added future efficiency in TAT and capacity in testing volumes. Methods We integrated a Thermo Scientific Transcend LX Multichannel (2-channel) UHPLC with our current Sciex ZenoTOF 7600 HRAM MS. Thermo Scientific provided the details for the contact closure cable, TMO link, and software needed. Thermo Scientific Aria MX was used for managing and controlling pump operations, XCalibur for batch building and Sciex OS for data acquisition and analysis. Residual clinical extracted runs previously analyzed and released on our single channel Agilent Infinity II UHPLC integrated with the Sciex ZenoTOF 7600 MS were reinjected onto a Sciex ZenoTOF 7600 MS using the 2-channel Thermo Scientific LC. Results The LC method for our quantitative risankizumab serum assay was translated from the Agilent to the Thermo LC without any changes to the gradient. We were successfully able to run across 2 channels using the 10-minute LC method with a 3-minute MS window to allow for multiplexing. Analytical sensitivity (measured as peak area counts) was comparable between systems. A N=22 patient comparison covering the measuring interval from 1 to 100 mcg/mL gave linear regression with a slope 0.971 and intercept of 0.7991; R2=0.9979. While the 2-channel LC could decrease the TAT in half, there is still the potential to implement a 3-channel LC with the 3-minute MS window from a 10-minute LC method. MRD is a qualitative assay with current LC run time of 13 minutes. Of the N=14 samples run for this comparison, there was 100% agreement for positivity and equivalent sensitivity compared to the single channel LC. By adding the 2-channel LC, we would be able to complete a 96-well plate for MRD in 12 hours instead of the current 24 hours. Conclusion We have successfully shown proof-of-concept for multiplexing a 2-channel UHPLC integrated with a HRAM TOFMS instrument, which has not been accomplished before. This will provide improved efficiency in turnaround time and open instrument capacity for future increases in testing volume without compromise in test cost, analytical sensitivity, or need for additional MS instrumentation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers