B-394 Advancing Rapid Clinical Toxicology Through DART-MS Detection of Emerging Drug Threats
Jill Kodger, Joe El-Khoury
Key points
From the paper's conclusion- Conclusion We developed a rapid DART-MS/MS method capable of detecting xylazine, medetomidine, and bromazepam in 10 seconds per sample.
Abstract
Background Novel psychoactive substances (NPS) are synthetic compounds engineered to mimic the effects of commonly misused illicit drugs. This study aims to develop a rapid and robust analytical method for detecting a novel benzodiazepine (bromazepam), as well as veterinary anesthetics such as xylazine and medetomidine. Interest in these substances has increased because of their growing involvement in polydrug exposures. At the same time, the continued spread of synthetic opioids, particularly fentanyl, has contributed to what is described as the “fourth wave” of the opioid epidemic. These evolving trends highlight the need for analytical platforms that can keep pace with rapidly changing drug use patterns. Current clinical laboratory technologies have not adapted quickly enough to address the emergence of these compounds. Immunoassays are slow to develop, and although liquid chromatography–tandem mass spectrometry (LC-MS/MS) offers high sensitivity, it is not widely accessible and often suffers from long turnaround times due to technical complexity. Direct analysis in real time (DART) tandem mass spectrometry (MS/MS) presents a promising alternative. DART is an ambient ionization technique that requires minimal sample preparation and eliminates the need for chromatographic separation, enabling near-instantaneous acquisition of mass spectral data. Methods Urine (50 µL) was combined with hydrolysis mix (50 µL) and left at room temperature for 5 minutes. Methanol (50 µL) was then added. Samples were vortexed for 10 seconds and centrifuged for 5 minutes at 15,000 × g at room temperature. A 5-µL aliquot of the supernatant was spotted onto the mesh of an Ion Sense DART QuickStrip and allowed to dry at room temperature before analysis. Results A 10-second jumpshot acquisition was used with a cone temperature of 250 °C and a DART gas temperature of 225 °C. The cone gas pressure was 23 psi. The analytical measureable range (AMR) was within %CV <12% for 1.3 to 2000 ng/mL for xylazine. The lower limit of quantitation was 5 ng/mL for xylazine and medetomidine, and 20 ng/mL for bromazepam. Conclusion We developed a rapid DART-MS/MS method capable of detecting xylazine, medetomidine, and bromazepam in 10 seconds per sample. With CLIA validation and application to clinical specimens, this approach could reduce turnaround times from days to seconds. Given the sensitivity and specificity of mass spectrometry, DART-MS/MS has the potential to serve as an alternative to traditional immunoassay-based drug screening.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical approaches for the identification and quantitation of nitazenes: a reviewJournal of Chromatography Open · 2025
- B-377 Rapid Evolution of the Fentanyl Drug Supply: Emergence of Medetomidine and Novel BenzodiazepinesClinical Chemistry · 2026
- Development and Validation of an Oral Fluid LC-MS/MS Method for Screening Emerging Synthetic Opioids Including Fentanyl Analogues, Nitazenes, and Xylazine.Journal of analytical toxicology · 2026
- Validation of a Dried Blood Spot (DBS) Liquid Chromatography - High-Resolution Mass Spectrometry (LC-HRMS) workflow for simultaneous identification of emerging opioids.Forensic science international · 2026
- 34 Development and Implementation of a Broad-spectrum Urine Drug Panel in a Clinical Toxicology LaboratoryAmerican Journal of Clinical Pathology · 2026
- 96-Sample Parallel BioSPME-DART-MS Platform for Rapid Screening of Fentanyl and its Analogues in Biofluids.Analytical chemistry · 2026