B-344 Enhanced Ultra Sensitive Measurement of Serum Estrogens Using the Xevo TQ Absolute XR with UHPLC MS/MS for Clinical Research Applications

Gareth Hammond, David Farrell, Robert Wardle, Lisa Calton, Margaret Guthrie, Michael Wakefield

Key points

From the paper's conclusion
  • These results support the use of this platform for low level estradiol measurement in clinical research..

Abstract

Background Accurate quantification of 17ß estradiol (E2) and estrone (E1) at low concentrations is essential for clinical research. Immunoassays often lack the analytical sensitivity and selectivity required at the low pg/mL range, while many LC MS/MS methods require large sample volumes or derivatization. The Xevo™ TQ Absolute XR Mass Spectrometer, incorporating extended dynamic range and an XR StepWave™ Ion Guide, offers enhanced analytical sensitivity and robustness that may improve low level estrogen quantification when paired with UHPLC separation and simplified sample preparation. Methods Certified reference materials for E2 and E1 and their 13C3-labeled internal standards were prepared in stripped human serum. Calibration ranges were 2–1000 pg/mL for E2 and 1–1000 pg/mL for E1. Quality control samples were prepared at 3.0, 10.0, and 700 pg/mL. A liquid–liquid extraction was performed on 250 µL of stripped serum (Golden West Biologicals, Temecula, CA) using hexane:ethyl acetate (85:15 v:v), followed by evaporation and reconstitution in methanol and water. Chromatographic separation was achieved on an ACQUITY™ UPLC™ I Class FL IVD System using a CORTECS™ Phenyl Column (2.1 x 50mm, 2.7µm) and a water/methanol/ammonium fluoride gradient. The Xevo TQ Absolute XR Mass Spectrometer operated in electrospray negative ionization mode with MRM monitoring for quantifier and qualifier transitions for each estrogen and its internal standard. Resolution was set to 0.7 FWHM for MS1 and MS2. Precision, linearity, analytical sensitivity, and matrix effects were evaluated across multiple occasions. Accuracy was assessed by comparison to UK NEQAS E2 proficiency materials. Results UHPLC MS/MS provided baseline chromatographic resolution of E2 and E1 with a total run time of 4.75 minutes. Calibration curves for both estrogens were linear across the full range, with correlation coefficients >0.999 and acceptable %bias across all five occasions. Within run and total precision were =3.2% CV at all QC concentrations for both analytes. Analytical sensitivity testing demonstrated that E1 could be reliably quantified at 1.0 pg/mL and E2 at 3.0 pg/mL, meeting criteria of =20% CV, =15% bias, and signal to noise =10:1. Distinction between blank serum and serum spiked at 1 pg/mL E2 was achievable without derivatization. Matrix factor studies across six individual sera indicated ion enhancement but was compensated for using the internal standard (E2 range -13.2% to 2.0% and E1 range -14.7% to 9.1%). Endogenous and exogenous interferences were assessed and the mean recovery observed was between 85–115%. Accuracy assessment showed strong agreement with UK NEQAS E2 values, with a Deming fit slope near 1.0 and a mean bias of -2.50%. Conclusion The Xevo TQ Absolute XR Mass Spectrometer, combined with UHPLC separation and simple liquid–liquid extraction, provides high analytical sensitivity quantification of serum estrogens within small sample volumes and without derivatization. The method demonstrated excellent linearity, precision, and agreement with external quality assessment materials. These results support the use of this platform for low level estradiol measurement in clinical research.. For Research Use Only. Not for use in diagnostic procedures.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers