B-339 Column Screening and SLE Extraction for LC/MS/MS Quantitation of Underivatized Estrone (E1) and Estradiol (E2) in Human Serum

Tim Fassette, Rajashree Chakravarti, Stephanie Marin

Key points

From the paper's conclusion
  • Conclusion Underivatized estrone and estradiol were quantitated by LC-MS/MS at low pg/mL concentrations using supported liquid extraction and a biphenyl column.

Abstract

Background Accurate quantification of estrone (E1) and estradiol (E2) is important for evaluating reproductive and endocrine function in clinical research. However, their inherently low circulating concentrations, limited ionization efficiency, and structural similarity present significant analytical challenges for direct liquid chromatography–tandem mass spectrometry (LC–MS/MS) without chemical derivatization. Although advancements in sample preparation and improvements in mass spectrometric sensitivity have enhanced assay performance, derivatization remains widely employed to achieve the sensitivity required for reliable detection in serum matrices. In this study, we describe the development of a highly sensitive LC–MS/MS method for the quantitation of underivatized E1 and E2 utilizing supported liquid extraction (SLE) and a biphenyl LC column. Methods Four reversed-phase LC columns were screened for optimal separation: Kinetex C18, Kinetex Biphenyl, Luna Omega Polar C18, and Kinetex F5. The Kinetex Biphenyl column (50 × 3.0 mm, 2.6 µm) provided the best resolution (∼1 min) and highest signal response for estrone. Samples were extracted using Strata-SE supported liquid extraction plates. 200µL of human serum were pre-treated with ammonium acetate buffer and extracted with 9:1 ethyl acetate:hexane. Extracts were dried under nitrogen and reconstituted in 60:40 ammonium fluoride:methanol. LC-MS/MS analysis was performed on an Agilent® 1290 Infinity system coupled with a SCIEX® 7500 TripleQuad mass spectrometer, using 0.5 mM ammonium fluoride in water and methanol with gradient elution over 5 minutes at 0.8 mL/min. Results Utilizing the pi-pi interactions available from a biphenyl LC column provided the best separation of structurally similar estrone and estradiol. The method demonstrated excellent linearity for E1 (1–500 pg/mL, R² = 0.998) and E2 (5–500 pg/mL, R² = 0.993). LOQs were 5 pg/mL for E1 and 10 pg/mL for E2, with signal-to-noise ratios of 12 and 8, respectively. Improved S/N for E2 was observed at 20 pg/mL (S/N = 19). It is reported in the literature that for E1, 95th percentile values in a population including males and females of all categories fall within 40-200 pg/mL and for E2 95th percentile values in a population of all males and females range from 40-350 pg/mL.1 This method using Strata SE SLE and Kinetex Biphenyl column (50 × 3.0 mm, 2.6 µm) on a high sensitivity MS provides a solution for the majority of samples and avoids time consuming, labor intensive derivatization typically used in estrogen methods. Conclusion Underivatized estrone and estradiol were quantitated by LC-MS/MS at low pg/mL concentrations using supported liquid extraction and a biphenyl column.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-339 Column Screening and SLE Extraction for LC/MS/MS Quantitation of Underivatized Estrone (E1) and Estradiol (E2) in Human Serum — Science Skim