B-169 Analytical performance verification of the automated NESCAUTO Cp Auto fecal calprotectin assay on the NS-Prime platform: assessment of comparability and reference intervals
Yoonjoo Kim, Su A Kim, Hyun young Ji, Seok Lae Chae
Key points
From the paper's conclusion- However, the significant proportional bias compared with the EliA Calprotectin assay on the Phadia 250 platform and the high positivity rate among healthy individuals at the manufacturer's cutoff underscore the need for establishing laboratory-specific reference intervals.
Abstract
Background Fecal calprotectin (FC) is a crucial non-invasive biomarker for the differential diagnosis and monitoring of inflammatory bowel disease (IBD). Although a fluorescence enzyme immunoassay (FEIA) provides automated results, it typically requires batch processing and has a longer turnaround time (TAT). The NESCAUTO Cp Auto NS-Prime (Alfresa Pharma, Japan) utilizes a colloidal gold agglutination method, allowing for rapid, random-access testing without manual sample pretreatment. This study aimed to verify the analytical performance of this rapid assay and evaluate its clinical comparability with an established FEIA. Methods Analytical verification was performed using the NS-Prime automated analyzer. Precision was evaluated according to CLSI EP05-A3 guidelines over 5 consecutive days (5 replicates per day) using low- and high-level control materials. Linearity was assessed according to CLSI EP06-A across five concentrations (24–1,264 µg/g) prepared by serially diluting calibrators. The manufacturer-claimed reference interval (RI) of <50 µg/g was verified per CLSI EP28-A3c using stool samples from 20 healthy individuals. For method comparison, 80 residual stool samples were analyzed using the NESCAUTO Cp Auto assay on the NS-Prime and the EliA Calprotectin immunoassay on the Phadia 250 (Thermo Fisher Scientific, Uppsala, Sweden). Bias estimation and method comparison were conducted according to CLSI EP09-A3. Results The assay demonstrated robust precision, with coefficients of variation (CV) of 1.3% (low) and 0.8% (high), surpassing manufacturer claims of 3.4% and 2.2%. Linearity was confirmed across the tested range (y=1.0095x-1.2219, R²=0.9988), with recoveries ranging from 90% to 110%. RI verification revealed that 75% (15/20) of healthy individuals exceeded the 50 µg/g cutoff value. Method comparison between the NESCAUTO and EliA assays showed a strong correlation (Spearman’s rho = 0.940, p<0.0001; Pearson's r = 0.891). However, categorical agreement at 50 µg/g was 82.5% (Kappa = 0.38). Passing-Bablok regression (slope = 0.835, intercept = 29.607) and Bland-Altman analysis indicated a systematic underestimation by the NESCAUTO assay (mean difference, -193.82 µg/g), particularly at higher concentrations. Conclusion The NESCAUTO Cp Auto NS-Prime provides excellent precision and linearity for routine FC measurement with a rapid TAT of 10–15 minutes. However, the significant proportional bias compared with the EliA Calprotectin assay on the Phadia 250 platform and the high positivity rate among healthy individuals at the manufacturer's cutoff underscore the need for establishing laboratory-specific reference intervals. Clinicians should be advised that numerical results are not directly interchangeable between these platforms when monitoring IBD activity, necessitating caution during longitudinal patient follow-up.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-280 Diagnostic Performance of Two Fecal Calprotectin Assays in an All-Comer Population Undergoing Evaluation for IBDClinical Chemistry · 2026
- Diagnostic prediction of ulcerative colitis using simplified endoscopic features and blood count-derived indices: a multicenter study.Frontiers in medicine · 2026
- Neutrophil percentage to albumin ratio and fecal calprotectin for distinguishing established inflammatory bowel disease from healthy controlsFrontiers in Medicine · 2026
- B-294 Retrospective review of the utility of faecal calprotectin at a large county safety net hospitalClinical Chemistry · 2026
- Fecal calprotectin concentrations in fecal extracts are associated with abnormal fecal status and gastrointestinal infection in captive and wild giant pandasBMC Veterinary Research · 2026
- B-217 Newly developed ELISA for detection of circulating calprotectin in serum and plasmaClinical Chemistry · 2026