B-053 Development of a Chemiluminescent Enzyme Immunoassay (CLEIA) for the Measurement of NT-proBNP

Nanami Tadokoro, Eri Inagaki, Yuichiro Miyazawa, Yoshiharu Makiguchi, Aya Oshimo, Miki Tanaka, Yuichi Ishii, Yoshiyuki Kitamura,

Key points

From the paper's conclusion
  • These results indicate that the assay provides reliable NT-proBNP measurement with good comparability to an established commercial assay, supporting its suitability for clinical use.

Abstract

Background Heart failure is a major cardiovascular disease with high prevalence and mortality, and N-terminal pro-B-type natriuretic peptide (NT-proBNP) is a well-established biomarker for its diagnosis and management. With changes in patient demographics and therapeutic strategies, accurate measurement of NT-proBNP has become increasingly important. Compatibility of NT-proBNP immunoassays across analytical platforms remains an important practical consideration for clinical laboratories. In February 2026, a newly developed NT-proBNP assay based on a chemiluminescent enzyme immunoassay (CLEIA) principle was introduced in Japan for use on the LUMIPULSE® L2400 system. Methods The analytical performance of the NT-proBNP assay using in-house-generated monoclonal antibodies was evaluated on the LUMIPULSE L2400 system in accordance with CLSI guidelines. The system provides two measurement modes: a normal turnaround time mode (NT mode) and a short turnaround time mode (ST mode). The reaction times were approximately 20 minutes for NT mode and 14 minutes for ST mode, resulting in total analytical turnaround times of approximately 32 minutes and 26 minutes, respectively. The limit of quantitation (LoQ) was determined in both NT and ST modes, and the reported LoQ was defined as the higher (worse) value observed between the two modes. Precision, linearity, and method comparison were assessed in both NT and ST modes using clinical samples covering clinically relevant concentration ranges. Interference and reference interval determination were performed in NT mode only. Method comparison was conducted using patient samples in comparison with an established commercial NT-proBNP immunoassay. Results The limit of quantitation (LoQ) of the NT-proBNP assay was 5.5 pg/mL (NT mode: 5.0 pg/mL; ST mode: 5.5 pg/mL). Within-laboratory precision showed coefficients of variation (CVs) of = 1.7-2.4 % in NT mode and = 1.8-3.0 % in ST mode across evaluated concentration levels. Linearity was demonstrated over a wide concentration range of 3.7-60795.4 pg/mL in NT mode and 3.0-55543.5 pg/mL in ST mode. No significant interference in NT mode was observed from unconjugated (=20.0 mg/dL) or conjugated bilirubin (=19.7 mg/dL), chyle (=1420 FTU), hemoglobin (=520 mg/dL), triglycerides (=2000 mg/dL), HAMA (=1237 ng/mL) or rheumatoid factor (=1149 IU/mL) at clinically relevant concentrations. Using samples from apparently healthy individuals (NT mode), the reference interval for serum NT-proBNP was determined to be 8.3-180.5 pg/mL, with a median value of 31.0 pg/mL in NT mode. Method comparison using patient samples demonstrated good agreement with an established commercial NT-proBNP immunoassay. The correlation coefficient (r) and regression slope were 1.0 and 1.09 in NT mode, and 1.0 and 1.12 in ST mode. (n = 161) Conclusion The NT-proBNP assay demonstrated a low limit of quantitation, a wide analytical measuring range, good linearity, and good correlation with an established commercial NT-proBNP immunoassay. The availability of ST mode with analytical performance comparable to that of the NT mode may support NT-proBNP testing in time-sensitive clinical settings. These results indicate that the assay provides reliable NT-proBNP measurement with good comparability to an established commercial assay, supporting its suitability for clinical use.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-053 Development of a Chemiluminescent Enzyme Immunoassay (CLEIA) for the Measurement of NT-proBNP — Science Skim