B-034 An end to subjective RPR reading? Development of a novel nontreponemal immunoassay for syphilis
Krista Meserve, Jose Cacicedo, Brittany Ahlstedt, Jennifer Petzke, Christopher Wisherd
Key points
From the paper's conclusion- Conclusion We have developed a novel nontreponemal immunoassay that consistently identifies moderately low to very high reactive and non-reactive syphilis samples and classifies antibody titer with no manual dilutions.
Abstract
Background Syphilis is a sexually transmitted infection caused by the Treponema pallidum bacterium. In 2023, 205,371 new cases of syphilis were reported in the United States, constituting an increase of 57.2% since 2019. Syphilis is diagnosed through a multi-test algorithm consisting of a treponemal test to detect specific antibodies produced against the bacteria, and a nontreponemal test to detect nonspecific reagin antibodies produced in response to cellular damage in the host. Nontreponemal testing centers around the rapid plasma reagin (RPR) method, a paper card-based manual assay marked by a lipid-antibody-charcoal agglutination reaction with a visual readout. This test is labor intensive and prone to subjectivity and intra- and inter-laboratory variability. Recent advances have included development of automated card imagers, but they rely on the same visual agglutination reaction. Our goal is to develop an automated, reproducible, and non-charcoal-based nontreponemal immunoassay that can be readily integrated into clinical lab spaces. Methods We have identified a novel method to attach a lipoidal antigen to a magnetic bead that allows the antigenic component of the nontreponemal assay to be integrated into an automated immunoassay with a turnaround time of under 30 minutes. Using human serum samples characterized as reactive (n=150), and non-reactive (n=100), we present a comparison of our novel immunoassay against traditional RPR cards (Arlington Scientific) and to an automated card reading platform (Gold Standard’s AIX1000). Additionally, we tested precision, linearity of dilution, and accelerated stability of our assay components. Results Our novel nontreponemal immunoassay has an overall agreement of 85.5% to the automated AIX1000 card-based system. Of reactive samples with moderately low to very high antibody levels (i.e. titers between 1:2 and 1:1024), 97% are reactive on our assay. Our assay performs equivalently to RPR card tests when reactive samples contain the lowest level of reagin antibodies (1:1 titer), which are the most subjective and difficult to identify samples using card-based methods. Non-reactive samples fall within a low, acceptable range on our assay. Based on relative signal of a single dilution, 91% of all reactive samples are classified at an acceptable dilution when compared to predicate titer. On average, intra- and inter-assay precision was less than 5% CV, samples diluted linearly and remained reactive after two weeks of accelerated stability testing. Ongoing work aims to improve recognition of reactive samples in the lowest titer zone. Conclusion We have developed a novel nontreponemal immunoassay that consistently identifies moderately low to very high reactive and non-reactive syphilis samples and classifies antibody titer with no manual dilutions. This fully automated immunoassay removes the need for highly trained technicians to read a visual output and can bring nontreponemal testing in-house, allowing for quicker treatment decision for patients suspected of having syphilis. This novel test dovetails well with our treponemal assay and can be performed simultaneously on the KleeYa platform, representing a significant advancement in automated syphilis diagnosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-160 Streamlining Syphilis Serology Through an Automated Multiplex Syphilis Total and RPR PanelClinical Chemistry · 2026
- Rabbit intradermal inoculations as an efficient method to quantitate Treponema pallidum.Journal of medical microbiology · 2026
- B-183 Evaluation of Synthetic Lipid Antigens as Alternatives to Traditional Purified Antigens in CSF VDRL TestingClinical Chemistry · 2026
- B-190 Increasing Syphilis Positivity and the Importance of Validated Rapid Testing for Early Diagnosis and Disease Control StrategiesClinical Chemistry · 2026
- Thrombin‐Antithrombin III Complex and Plasmin‐α2‐Plasmininhibitor Complex as Novel Biomarkers for Neurosyphilis: A Pilot StudyInternational Journal of Clinical Practice · 2026
- Stage-Specific Associations of Automated RPR With Treponema pallidum qPCR Positivity and Cycle Threshold in Early Symptomatic Syphilis: A Multicenter Study Using Specimens Collected During Routine CaremedRxiv · 2026