Clinical Chemistry

A-341 Analytical Performance Verification of sFlt-1 and Placental Growth Factor Immunoassays for Pre-eclampsia

Mary Monyae

Observational

Key points

From the paper's conclusion
  • Conclusion The Siemens Atellica IM platform demonstrated robust analytical performance for PlGF and sFlt-1, supporting its suitability for routine laboratory use.

Abstract

Background Placental growth factor (PlGF) and soluble fms-like tyrosine kinase-1 (sFlt-1) are angiogenic biomarkers used to support the diagnosis and risk stratification of pre-eclampsia. International recommendations for their clinical use are assay-specific and largely based on manufacturer-validated platforms. Limited data is available on the analytical performance and inter-platform comparability of these biomarkers in the South African laboratory setting, particularly across central laboratory and point-of-care systems. Methods This retrospective laboratory-based study was conducted at a tertiary National Health Laboratory Service facility in Johannesburg, South Africa. Analytical verification was performed in accordance with CLSI EP15, EP06, and EP09 guidelines. PlGF was measured on the Siemens Atellica IM, Thermo Fisher BRAHMS Kryptor, and Aucheer i-RAte platforms, while sFlt-1 was assessed on the Atellica IM and i-RAte systems. Precision was evaluated using manufacturer-provided and third-party quality control materials. Linearity was assessed on the Atellica IM platform using serial dilutions of calibration material. Method comparison included 64 residual serum samples analysed using Passing–Bablok regression and Bland–Altman analysis. Diagnostic concordance was assessed using an sFlt-1 to PlGF ratio cutoff of less than or equal to 38. Results The Atellica IM platform demonstrated acceptable analytical precision for both PlGF and sFlt-1, with within-run and between-day coefficients of variation below 6 percent across quality control levels. Linearity on the Atellica IM was confirmed across clinically relevant ranges for PlGF (9 to 1301 pg/mL) and sFlt-1 (15 to 17 595 pg/mL), with all deviations remaining within the allowable systematic error of ±12.5 percent. Method comparison showed strong correlation between Atellica IM and Kryptor for PlGF (?=0.92), with a mean bias of -9.1 percent. Comparisons involving the point-of-care platform demonstrated weaker correlations and substantial positive bias exceeding 60 percent for both PlGF and sFlt-1. Concordance analysis of the sFlt-1 to PlGF ratio between Atellica IM and the point-of-care platform demonstrated an overall agreement of 82.2 percent with moderate agreement by Cohen’s kappa (?=0.62). Conclusion The Siemens Atellica IM platform demonstrated robust analytical performance for PlGF and sFlt-1, supporting its suitability for routine laboratory use. Significant inter-platform bias, particularly involving the point-of-care system, indicates that angiogenic biomarker results should not be used interchangeably across platforms. These findings highlight the need for assay-specific cutoffs and local analytical verification prior to clinical implementation, especially in resource-limited settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers