A-339 Performance Evaluation of New sFlt-1 and PlGF Immunoassays on Mindray CL-6000i
Wenyuan Lu, Yeng Xiong, Terri McLernon, Sarah Barclay, Obsa Bokku
Key points
From the paper's conclusion- This study evaluates the performance of two new immunoassays for the measurements of sFlt-1 and PlGF on Mindray CL-6000i.
Abstract
Background Preeclampsia is one of the leading causes of preterm delivery. According to the WHO, preeclampsia affects 3%-8% of pregnancies, leading to approximately 42,000 maternal deaths and much higher pre-term infant deaths annually [1]. In Asia, Africa, and Latin America, preeclampsia accounts for as many as one-quarter of maternal deaths. The severity of the disease and healthcare expenses have created a strong need for early predictive tools to identify preeclampsia and provide appropriate care. Academic and industrial research has heavily focused on biomarker-based detection, particularly the sFlt-1/PlGF ratio, and molecular-level studies for improved risk stratification. This study evaluates the performance of two new immunoassays for the measurements of sFlt-1 and PlGF on Mindray CL-6000i. Methods We coupled tosyl-activated superparamagnetic beads (Dynabeads™) with anti-human sFlt-1 and PlGF capture antibodies. Alkaline phosphatase conjugates of anti- human sFlt-1 and PlGF antibodies were paired with each assay’s capture antibody. Three-level calibration and three level controls in liquid form were created that can be stored at 4°C-8°C. Precision, linearity, hook limit, onboard stability, interference from cross reactants, and correlation were evaluated on Mindray’s automated chemiluminescent analyzer, CL-6000i. Roche Diagnostics Elecsys sFlt-1 and PlGF were used to assign calibrator values. Results Precision was run with three in-house controls and three serum pools. Repeatability (20 replicates) ranged from 0.7% to 2.5% for sFlt-1 (24-55,000pg/ml) and 1.0% to 5.1% for PlGF (5-3700pg/ml). Within-lab precision (20 days of repeats) was satisfactory (sFlt-1: 1.6% to 4.6% at 36-56,000pg/ml, SD 1.2pg/ml at 25.6pg/ml. PlGF: 2.4% to 6.5% at 80-3,700pg/ml, SD 0.5pg/ml at 5.5pg/ml). LoB, LoD, and LoQ were evaluated using five samples tested in quadruplicate over three days. Values were 1.7pg/mL, 2.2pg/mL, and 11.6pg/mL for sFlt-1 and 0.9pg/mL, 2.1pg/mL, and 8.3pg/mL for PlGF, respectively. LoQ was determined based on the calculation for functional sensitivity (CV=20%). Upper measuring range was 85,000pg/ml for sFlt-1 and 10,000pg/ml for PlGF. No hook limit was observed up to 200,000pg/ml for sFlt- and 100,000pg/ml for PlGF. Reagent onboard stability was demonstrated for 28 days without recalibration for both assays. Accelerated stability study was done at 37°C for 12 days. Data from three in-house controls and three serum pools demonstrated reagent pack dose drift ranged from -5% to 4% for sFlt-1 and -5% to 7% for PlGF. Two calibrators revealed drift of -6.6% at 180pg/ml sFlt-1, -3.2% at 9,540pg/ml sFlt-1, -8.7% at 210pg/ml PlGF, and -9.5% at 3,720pg/ml PlGF. The assays correlated well with Roche Elecsys kits (sFlt-1: Y=1.04X+5.6, r-squared=0.985, n=63, linear regression. PlGF: Y=0.99X-3.3, r-squared=0.983, linear regression. Ratio sFlt-1/PlGF: Y=0.90X+0.22, Passing-Bablok regression). Cross-reactivities to VEGF165, VEGF167, PlGF2, and heterodimer VEGF/PlGF1 were comparable with predicate kits. Conclusion From this evaluation, we concluded that the prototype sFlt-1 and PlGF assays can measure serum sFlt-1 and PlGF precisely and accurately with good performance, and are suitable for further evaluation in clinical settings and ultimately for use in routine clinical laboratories. Reference: [1] www.WHO.int, “Pre-eclampsia”, December 10, 2025. Acknowledgement: We thank Jie Zhang, Lei Cao, and Pingzhang Chen for their contributions to this study at Mindray, Shenzhen, China.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-341 Analytical Performance Verification of sFlt-1 and Placental Growth Factor Immunoassays for Pre-eclampsiaClinical Chemistry · 2026
- A-340 Clinical Implementation of sFlt-1/PlGF Biomarker Testing for Preeclampsia Risk Assessment at a Tertiary Care Centre: Summary of 5-year ExperienceClinical Chemistry · 2026
- sFlt-1/PlGF Ratio as a Central Biomarker for Preeclampsia and Perinatal Outcomes: A Multisystem Retrospective Cohort StudyJournal of Clinical Medicine · 2026
- Emerging biomarkers and integrated models for preeclampsia prediction.American journal of clinical and experimental immunology · 2026
- Present status of sFLT1 and PlGF as diagnostic and therapeutic targets for preeclampsiaThe Journal of Maternal-Fetal & Neonatal Medicine · 2026
- Predictive Value of the Soluble fms-Like Tyrosine Kinase-1/Placental Growth Factor Ratio in Preeclampsia and Placenta Accreta Spectrum: A Prospective Observational Study.Cureus · 2026