A-216 510(k) clearance of the universally applicable DiaSys Procalcitonin FS test for the diagnosis of sepsis

Leoni Wey, Thomas Masetto, Christian Kochem, Tim Fleissner, Zeynep Yetiz, Dagmar Schmidt, Kai Prager, Tabea Rosenbaum,

Key points

From the paper's conclusion
  • Conclusion The PCT FS assay received FDA 510(k) clearance on May 9, 2025 (K242294), enabling its use as an aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

Abstract

Background Since 2020, DiaSys has marketed an immunoturbidimetric assay for the quantification of procalcitonin (PCT) in human serum and plasma in Europe under CE marking. To support U.S. market authorization, the U.S. Food and Drug Administration (FDA) requested additional technical data for 510(k) clearance, including evaluations of human anti-mouse antibody (HAMA) interference, total protein interference, sample stability, Cohen’s kappa at clinically relevant cut-off values, and negative and positive percent agreement with a reference method. Methods All evaluations were performed in human serum and plasma using the Abbott ARCHITECT c8000 as the predicate device. An acceptable recovery range of ±15% was predefined. HAMA interference was assessed at concentrations up to 480 ng/mL by spiking highly concentrated HAMA material into human samples containing approximately 0.5 µg/L PCT. Total protein interference was evaluated at concentrations up to 139 g/L by spiking bovine serum albumin (BSA) into a human sample. For sample stability testing, fifteen human serum and plasma samples were spiked with PCT concentrations covering the full analytical measurement range. Aliquots were stored for 24 hours at 20–25 °C, 5 days at 2–8 °C, and 14 days at -20 °C. Stability was assessed using linear regression analysis and scatter plots. Methods comparison was conducted using 210 human samples analyzed with the PCT FS assay and the bioMérieux VIDAS® Procalcitonin reference method. Negative percent agreement (NPA) and positive percent agreement (PPA) were calculated using 2 × 2 contingency tables at clinical cut-off values of 0.5 and 2.0 µg/L. Cohen’s kappa was calculated using MedCalc® Statistical Software (version 23.1.7). Acceptance criteria for NPA and PPA were ±10%. Results The immunoturbidimetric PCT FS assay showed no evidence of interference from HAMA. Total protein concentrations up to 139 g/L did not affect assay recovery. Human serum and plasma samples were stable under all tested temperature conditions for the specified storage durations. Compared with the reference method, the PCT FS assay demonstrated a Cohen’s kappa of 0.981 (95% CI: 0.954–1.000) at the 0.5 µg/L cut-off, with a negative percent agreement (NPA) of 98.0% (95% CI: 92.9%–99.8%) and a positive percent agreement (PPA) of 100% (95% CI: 96.7%–100.0%). At the 2.0 µg/L cut-off Cohen’s kappa was determined with 0.955 (95% CI: 0.912–0.999), with a negative percent agreement (NPA) of 100.0% (95% CI: 97.5%–100.0%) and a positive percent agreement (PPA) of 94.0% (95% CI: 85.4%–98.4%). Conclusion The PCT FS assay received FDA 510(k) clearance on May 9, 2025 (K242294), enabling its use as an aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers