Clinical Chemistry

A-199 High Diagnostic Accuracy of an Improved Fully Automated IGRA (LIAISON® QuantiFERON TB Gold Plus II) for Active TB and Low Risk Screening

Greg Feitl, Ashli Rode, Lezlie Kelzenberg, Luca Pallavicini, Dat Vo, Francesco Capuano

Key points

From the paper's conclusion
  • Results The improved LIAISON® QFT II assay demonstrated high diagnostic accuracy and robust signal separation around the 0.35 IU/mL cutoff, supporting confident interpretation across risk groups.Active TB sensitivity: 86.9% (119/137; 95% CI: 80.2–91.5%)Low risk specificity: 98.3% (170/173; 95% CI: 95.0–99.4%) Distribution analysis revealed distinct clustering of true negatives well below the cutof

Abstract

Background Accurate identification of Mycobacterium tuberculosis (TB) infection is essential for early treatment and prevention strategies. Interferon gamma release assays (IGRAs) play a key role in this effort, but must deliver strong diagnostic accuracy, clear positive–negative separation, and high reliability across patient populations. The improved LIAISON® QuantiFERON TB Gold Plus II (QFT II) assay was redesigned with enhanced antibody and detection chemistry to strengthen signal performance at the decision cutoff while maintaining seamless laboratory automation. Methods A total of 774 subjects were prospectively evaluated across clinically distinct cohorts including culture confirmed active TB (n=146), low risk individuals (n=173), and routine screening populations (n=455). Plasma samples were processed on the fully automated LIAISON® platform. Diagnostic sensitivity in active TB and specificity in low risk individuals were calculated. Cutoff behavior was assessed through distribution analysis of TB1 Nil and TB2 Nil values. Results The improved LIAISON® QFT II assay demonstrated high diagnostic accuracy and robust signal separation around the 0.35 IU/mL cutoff, supporting confident interpretation across risk groups.Active TB sensitivity: 86.9% (119/137; 95% CI: 80.2–91.5%)Low risk specificity: 98.3% (170/173; 95% CI: 95.0–99.4%) Distribution analysis revealed distinct clustering of true negatives well below the cutoff and strong positive responses in active TB, minimizing borderline results and supporting reliable discrimination. Performance remained strong in challenging populations such as pediatrics, immunocompromised subjects, and pregnant women, with high percent agreement and minimal indeterminate outcomes. Conclusion The improved LIAISON® QFT II assay delivers high sensitivity and specificity with excellent positive–negative separation at the diagnostic cutoff, enabling confident IGRA interpretation across diverse clinical settings. Combined with full automation, reproducible signal performance, and streamlined workflow, QFT II supports high throughput TB screening and reliable detection of active TB infection - strengthening both laboratory efficiency and public health impact.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers