A-152 Establishment of fasting C-peptide reference interval by Roche e 801 in a Singaporean population

Amas Lee, Bobby Li, Hock Ang Teo, Robert Hawkins

Key points

From the paper's conclusion
  • Both the upper and lower reference limit in our population was significantly lower than the reference interval provided by the manufacturer in the kit insert (370 – 1470 nmol/L), demonstrating the need to tailor the C-peptide reference interval to the local population.

Abstract

Background C-peptide measurements are useful in the investigation of hypoglycaemia and in the assessment of the ß-islet cell reserve in diabetes mellitus. The aim of the study was to establish a central 95% reference interval (RI) for C-peptide for the Singaporean population. Methods Leftover serum from fasting healthy adults presenting for health screening at an outpatient Heath Enrichment Clinic. Exclusion criteria were current or previous HbA1c >6.5% or glucose >5.5 mmol/L. All C-peptide measurements were performed on a Roche cobas e801 instrument (Roche Diagnostics, Mannheim, Germany) using the Elecsys C-Peptide kit. Results 292 adults (148 females, 144 males) were included. The C-peptide concentrations did not follow a Gaussian distribution and required log-transformation. The log-transformed parametric central 95th percentile reference interval (with 90% confidence interval of reference limits) for males and females were 320 (297 – 346) to 1143 (1060 – 1233) pmol/L and 286 (265 – 308) to 1143 (972 – 1133) pmol/L respectively. The non-parametric central 95th percentile reference interval (with 90% confidence interval of reference limits) for males, females and overall were 308 (251 – 355) to 1216 (1034 – 1303) pmol/L, 265 (244 – 331) to 1090 (1045 – 1306) pmol/L, and 298 (256 – 328) to 1130 (1048 – 1256) pmol/L respectively. Conclusion There was a statistically significant difference between the median C-peptide for men and women, but the difference in the upper and lower reference interval were not clinically significant enough to justify gender-specific reference interval. The non-parametric mixed gender reference interval of 298 – 1130 pmol/L was chosen for use in the Singaporean population was chosen. Both the upper and lower reference limit in our population was significantly lower than the reference interval provided by the manufacturer in the kit insert (370 – 1470 nmol/L), demonstrating the need to tailor the C-peptide reference interval to the local population.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-152 Establishment of fasting C-peptide reference interval by Roche e 801 in a Singaporean population — Science Skim