A-152 Establishment of fasting C-peptide reference interval by Roche e 801 in a Singaporean population
Amas Lee, Bobby Li, Hock Ang Teo, Robert Hawkins
Key points
From the paper's conclusion- Both the upper and lower reference limit in our population was significantly lower than the reference interval provided by the manufacturer in the kit insert (370 – 1470 nmol/L), demonstrating the need to tailor the C-peptide reference interval to the local population.
Abstract
Background C-peptide measurements are useful in the investigation of hypoglycaemia and in the assessment of the ß-islet cell reserve in diabetes mellitus. The aim of the study was to establish a central 95% reference interval (RI) for C-peptide for the Singaporean population. Methods Leftover serum from fasting healthy adults presenting for health screening at an outpatient Heath Enrichment Clinic. Exclusion criteria were current or previous HbA1c >6.5% or glucose >5.5 mmol/L. All C-peptide measurements were performed on a Roche cobas e801 instrument (Roche Diagnostics, Mannheim, Germany) using the Elecsys C-Peptide kit. Results 292 adults (148 females, 144 males) were included. The C-peptide concentrations did not follow a Gaussian distribution and required log-transformation. The log-transformed parametric central 95th percentile reference interval (with 90% confidence interval of reference limits) for males and females were 320 (297 – 346) to 1143 (1060 – 1233) pmol/L and 286 (265 – 308) to 1143 (972 – 1133) pmol/L respectively. The non-parametric central 95th percentile reference interval (with 90% confidence interval of reference limits) for males, females and overall were 308 (251 – 355) to 1216 (1034 – 1303) pmol/L, 265 (244 – 331) to 1090 (1045 – 1306) pmol/L, and 298 (256 – 328) to 1130 (1048 – 1256) pmol/L respectively. Conclusion There was a statistically significant difference between the median C-peptide for men and women, but the difference in the upper and lower reference interval were not clinically significant enough to justify gender-specific reference interval. The non-parametric mixed gender reference interval of 298 – 1130 pmol/L was chosen for use in the Singaporean population was chosen. Both the upper and lower reference limit in our population was significantly lower than the reference interval provided by the manufacturer in the kit insert (370 – 1470 nmol/L), demonstrating the need to tailor the C-peptide reference interval to the local population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Commonly used C‐peptide assays show differing associations with CGM metricsDiabetes Obesity and Metabolism · 2025
- Glycaemic Variability and C‐Peptide in Bangladeshi Children, Adolescents, and Young Adults With Diabetes: A Prospective StudyEndocrinology, Diabetes & Metabolism · 2026
- The Relationship Between Smoking Status and C-Peptide Levels in Patients with Type 2 Diabetes Mellitus.Metabolites · 2026
- Low Serum C-Peptide and Albuminuria Severity in Type 2 Diabetes Mellitus: A Cross-Sectional Study.Journal of clinical medicine · 2026
- Utilidad Clínica del Péptido C en el Manejo de la Diabetes Tipo 2 Insulinorrequiriente: Revisión SistemáticaCiencia Latina Revista Científica Multidisciplinar · 2026
- ANTI-GAD VÀ C-PEPTIDE Ở BỆNH NHÂN ĐÁI THÁO ĐƯỜNG TYPE 2 KHỞI PHÁT SỚM KHÔNG THỪA CÂN BÉO PHÌ: GỢI Ý TẦM SOÁT LADATạp chí Y học Cộng đồng · 2026