A-147 Performance of an IFCC-Compliant Immunoturbidimetric HbA1c Assay: A Study of Consistency with the Gold Standard
Chengdong Ji, Yun Li, Qian Chen, Y Yang
Key points
From the paper's conclusion- This study develops an oligomeric antibody-based immunoturbidimetric method to achieve HPLC-comparable accuracy with enhanced stability, offering a practical clinical alternative.
Abstract
Background Glycated hemoglobin (HbA1c) reflects average blood glucose over 2–3 months and is the gold standard for diabetes diagnosis and monitoring, as recommended by the American Diabetes Association. High-performance liquid chromatography (HPLC) is the gold standard for HbA1c but is slow and costly. Traditional immunoturbidimetric assays suffer from complexity and instability. This study develops an oligomeric antibody-based immunoturbidimetric method to achieve HPLC-comparable accuracy with enhanced stability, offering a practical clinical alternative. Methods This study evaluated the Zybio HbA1c immunoturbidimetric kit (oligomeric antibody technology) against the HPLC gold standard (Arkray HA-8180) using 194 clinical samples. Regression analysis assessed consistency, and analytical performance was compared with traditional immunoturbidimetric assays. Results An immunoturbidimetric assay for HbA1c detection was constructed using oligomerized anti-glycated hemoglobin antibodies (Figure 1A). The consistency between the Zybio’s HbA1c kit and HPLC an R2 value of 0.9928, as shown in Figure 1B.Within the linear range of?[3.6%,14%][3.6%,14%], the regression equatio was?Y=1.0118X+0.1999Y=1.0118X+0.1999, with a correlation coefficient?r=0.9979r=0.9979, exceeding the required threshold of 0.975,(Figure 1C,D). HbA1c kit eliminated the polyclonal antibody component used in traditional immunoturbidimetric assays, leading to a significant performance improvement over conventional assay method. The traditional method exhibited a low-end repeatability CV = 7%, whereas HbA1c kit demonstrated a repeatability CV < 2.5%(Figure 1E,F). Moreover, the relative deviation of Zybio’s HbA1c reagent after 30 days of in-use stability was < 5%, while the deviation for the traditional method exceeded 7% after just 15 days of in-use stability(Figure 1G,H). Conclusion This study applies oligomeric antibody technology to improve the immunoturbidimetric detection of HbA1c, addressing the stability and repeatability limitations of traditional methods. Additionally, in clinical practice, the detection results of Zybio’s HbA1c reagent kit demonstrated excellent consistency with those obtained by HPLC.The detection results of Zybio’s HbA1c kit have also received continuous certification from the IFCC. The evaluated HbA1c kit demonstrates reliable performance in clinical agreement, stability, and repeatability, making it suitable for routine diagnostics. Its automated, high-speed platform effectively meets clinical needs for accuracy, efficiency, and cost-effectiveness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative analysis of enzymatic assay for HbA1c measurement with high performance liquid chromatography.Scientific reports · 2026
- A-136 Improving HbA1c Result Reportability Using Dual-Method Verification: Bio-Rad D-100 HPLC and Siemens Healthineers DCA ImmunoassayClinical Chemistry · 2026
- Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method.Diabetology international · 2026
- B-059 Precision and Bias Verification of HbA1c by HPLC Following CLSI EP15-A3 in a Hospital Laboratory in Manabí, EcuadorClinical Chemistry · 2026
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- Comparison of Two HbA1c Measurement Methods and Their Analytical Performance: TINIA vs. Boronate Affinity HPLC.Clinical laboratory · 2026