Clinical Chemistry

A-101 Comparison of BioFire and Verigene Blood Culture Tests for Identification of Polymicrobial Gram-Positive/Gram-Negative Cultures

George Pratt, Sierra Amaya, Jose Gomez-Garcia, Victor Kubatin, Nadezhda Kubatin, Lokinendi Rao

Key points

From the paper's conclusion
  • This study compared the performance of the BioFire Blood Culture Identification 2 (BCID2) Panel and Verigene Blood Culture Nucleic Acid Tests against traditional mass spectrometry (MALDI-TOF) and antimicrobial susceptibility testing (AST) in polymicrobial blood cultures.

Abstract

Background Rapid detection and identification of pathogenic organisms in blood cultures is critical for managing sepsis patients. Molecular diagnostic panels have recently emerged to identify pathogens and genetic markers of antimicrobial resistance in positive blood cultures, providing results within hours rather than days after growth detection. This study compared the performance of the BioFire Blood Culture Identification 2 (BCID2) Panel and Verigene Blood Culture Nucleic Acid Tests against traditional mass spectrometry (MALDI-TOF) and antimicrobial susceptibility testing (AST) in polymicrobial blood cultures. Methods Fifteen consecutive polymicrobial blood culture samples containing both gram-positive and gram-negative organisms, routinely submitted to Quest Diagnostics, were anonymized and analyzed using both molecular and microbiological methods. Following Gram staining, samples were tested on the Verigene and BioFire platforms and subsequently analyzed by MALDI-TOF mass spectrometry and the VITEK 2 system for pathogen identification and AST. Results from the molecular panels were compared with each other and against MALDI-TOF/VITEK 2 as the reference standard. Results Overall, the BioFire and Verigene molecular panels demonstrated high concordance. Of the 15 samples tested, 13 had pathogens successfully identified by the molecular assays. MALDI-TOF/VITEK 2 and phenotypic susceptibility testing detected organisms and resistance mechanisms not targeted by the molecular panels, accounting for the 2 samples that yielded no molecular identification. Compared with MALDI-TOF/VITEK 2, BioFire detected more organisms than Verigene and identified 3 Enterococcus faecalis isolates that MALDI-TOF failed to detect. Conclusion This comparative analysis demonstrates that both BioFire BCID2 and Verigene molecular panels offer reliable rapid pathogen identification in polymicrobial blood cultures, with high concordance between platforms for Gram-negative organisms. BioFire exhibited broader detection capability for Gram-positive organisms, including identification of Enterococcus faecalis in 3 samples undetected by MALDI-TOF/VITEK 2. While it remains uncertain whether these discrepancies reflect false-positive molecular results or superior analytical sensitivity, they underscore the complementary value of molecular and traditional microbiological methods. The molecular panels’ rapid turnaround time provides critical early diagnostic information, while MALDI-TOF/VITEK 2 retains essential utility for detecting organisms beyond the molecular panel targets and performing comprehensive antimicrobial susceptibility testing. Future studies with larger sample sizes and clinical outcome data are warranted to fully elucidate the impact of these molecular diagnostic platforms on antimicrobial stewardship and patient outcomes in sepsis management.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers