Density measurement results of the Cotton–Mouton effect polarimetry on EAST
Minyong Shen, Jibo Zhang, Zou zhiyong, Nu Zhang, Yao Zhang, Yinxian Jie, Haiqing Liu, Jinlin Xie,
Key points
From the paper's conclusion- At a wavelength of 432.6 µm, a toroidal magnetic field of BT = 2.35T, and a chord length of 1.35 m, the expected CM phase shift along the EAST central chord is approximately 8.5◦ for an electron density of 1 × 1019 m−3.
Abstract
Conventional interferometers may lose the integer number of 2π fringes after signal interruption, limiting the reliability of electron-density measurements during long-pulse plasma operation. To address this problem, a far-infrared (FIR) polarimeter–interferometer based on the Cotton–Mouton (CM) effect was developed, and plasma-induced CM phase shifts were measured for the first time on the Experimental Advanced Superconducting Tokamak (EAST). The system measures the birefringent differential phase shift between two orthogonal polarization components, which is approximately proportional to ∫neB⊥2 dl. Statistical analysis of multiple discharges shows an approximately linear relationship between the CM phase shift and the line-integrated electron density, consistent with CM theory. At a wavelength of 432.6 µm, a toroidal magnetic field of BT = 2.35T, and a chord length of 1.35 m, the expected CM phase shift along the EAST central chord is approximately 8.5◦ for an electron density of 1 × 1019 m−3. The system also resolves MHD modes in the kilohertz range, exhibits a long-term baseline drift of only 0.02◦, and maintains the intermediate frequency (IF) within 850±50 kHz during consecutive discharges through CO2 laser frequency stabilization. The CM phase can be recovered directly after signal interruption without reconstructing the accumulated phase history, providing an independent reference for identifying and correcting 2π fringes jump in conventional interferometric measurements. This work establishes a technical basis for reliable long-pulse electron-density measurement and feedback control on EAST and future fusion devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multifunctional interferometry-polarimetry for fusion plasma control and physics applications (invited).The Review of scientific instruments · 2026
- Simultaneous full- and half-energy Stark polarization measurements using a modified background polychromator motional Stark effect system.The Review of scientific instruments · 2026
- Integrated frequency-modulated continuous-wave (FMCW) interferometer with in situ common-path referencing for plasma diagnostics using a commercial radar system-on-chip.The Review of scientific instruments · 2026
- Density profile reconstruction and fringe jump correction through the combination of limited-chord toroidal interferometer and polarimeter.The Review of scientific instruments · 2026
- Generalized theory of second-harmonic orthogonally polarized dispersion interferometry and implementation for pulsed power electrode plasma measurements.The Review of scientific instruments · 2026
- Development and validation of a multichannel edge density profile reflectometer embedded in ion cyclotron resonance heating (ICRH) antennas on EAST.The Review of scientific instruments · 2026