Autocorrelation-based signal quality assessment for ballistocardiogram: a peak-free approach.
Huilin Ren, Sining Li, Dongjun Zhang, Qianghua Liao, Deng Minzhi, Dan Wu, Bochao Su
Shenzhen Polytechnic University, 7098 Xueyuan Avenue, Nanshan District, Shenzhen, 518055, China. · Tech X Academy, Shenzhen Polytechnic University, 7098 Xueyuan Avenue, Nanshan District, Shenzhen, 518055, China.
Key points
From the paper's conclusion- We propose a peak-free SQA method that evaluates signal periodicity through windowed autocorrelation, retains power spectral concentration only as an auxiliary descriptor, and aggregates the autocorrelation peaks by the median into a robust subject-level index.
Abstract
Ballistocardiogram (BCG) signal quality assessment (SQA) is critical for reliable cardiac monitoring and disease prediction. Existing SQA methods for physiological signals typically rely on peak detection, which is unreliable for BCG due to its multi-lobed morphology. We propose a peak-free SQA method that evaluates signal periodicity through windowed autocorrelation, retains power spectral concentration only as an auxiliary descriptor, and aggregates the autocorrelation peaks by the median into a robust subject-level index. Evaluated on a 46-subject dataset with synchronized reference ECG, the method yields an interpretable quality stratification. Using the reference ECG as ground truth, the index is strongly predictive of BCG usefulness: high-quality windows achieve substantially lower heart-rate error and a much higher proportion of clinically usable segments compared to low-quality windows. In discriminating usable windows, the proposed index significantly outperforms peak-based and statistical SQI baselines. Furthermore, the screened 1D-ConvVAE pipeline achieved an AUC of 0.815 versus 0.675 for the full cohort. A same-model, within-cohort stratified evaluation showed higher AUC on good- than lower-quality subjects (0.765 vs.\ 0.587); a filter ablation indicates that bandpass filtering alone does not account for the observed difference. The method provides a reliable, peak-free SQA approach for BCG, enabling automatic quality screening for large-scale datasets.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Dual-Channel Signal Quality Assessment Framework for Reliable Seated Ballistocardiographic Biosensing with Measurement Configuration Optimization.Biosensors · 2026
- Global-key region collaborative noise-aware for electrocardiogram signal quality assessment.PloS one · 2026
- Real-time AI-based signal quality control enhances signal reliability and diagnostic stability in multi-site digital auscultationEuropean Heart Journal · 2026
- A Novel K-Means with SHAP Feature Selection and ROA-Optimized SVM for Sleep Monitoring from Ballistocardiogram SignalsMathematics · 2026
- The autonomic-ventricular coherence index: a novel single-lead ECG method with acquisition-dependent discrimination.Frontiers in digital health · 2026
- Artificial intelligence driven software systems for cardiovascular disease detection using physiological signals: a systematic reviewArtificial Intelligence Review · 2026