Shear Rate and Temperature Dependent Characterization of Blood-Mimicking Fluids for Blood Warming Device Application.
Andrea Riddle, Emine Foust
United States Military Academy at West Point, Mahan Hall, Bldg 752, West Point, West Point, New York, 10996, United States.
Key points
From the paper's conclusion- At shear rates ≥ 100 s⁻¹, 50% glycerol-water solutions produced viscosity values closest to reported whole blood values.
Abstract
During emergency transfusion, stored whole blood at 1-6°C must be rapidly warmed to 36-38°C using blood warming devices. However, device validation commonly relies on Newtonian crystalloid intravenous fluids with a viscosity of 1 mPa·s, whereas whole blood exhibits temperature-dependent, shear-thinning rheology with viscosities of 3.5-4.5 mPa·s at 37°C. This discrepancy introduces systematic errors in performance evaluation. To address this gap, this research study characterized three blood-mimicking fluids (BMFs), including glycerol-water solutions of 22-50% w/w, egg whites, and TrueClot across temperatures of 5-40°C and shear rates of 5-100 s⁻¹. Among the tested fluids, TrueClot most closely resembled blood in overall rheological properties. However, TrueClot demonstrated abnormal phase separation at temperatures below 10°C, with a 28% drop in viscosity from 55 ± 4 to 39.6 ± 0.5 mPa·s over just 2°C, limiting its use at refrigerated storage temperatures. At shear rates ≥ 100 s⁻¹, 50% glycerol-water solutions produced viscosity values closest to reported whole blood values. Based on the rheological profiles obtained, it is inferred that at shear rates < 100 s⁻¹, diluted TrueClot at temperatures higher than 10°C or modified egg whites with xanthan gum at 5-40°C may provide suitable shear-thinning properties, though experimental validation of these modified solutions is required. Two mathematical models, including the Casson model with Apostolidis and Beris parametrization and a combined shear-temperature power law model, were developed to predict viscosity values that correlate well with the experimental results.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Geometry-Dependent Variability in Rotational Rheometry of Whole Blood: Implications for Flow Curve Measurement and Protocol Standardization.Methods and protocols · 2026
- A synergistic framework of homotopy perturbation and finite element analysis of electro-osmotic thermo-hydrodynamics for Carreau-Yasuda hybrid nanofluids in composite arterial stenoses.Computational biology and chemistry · 2026
- Blood rheology-based analytical framework for time-dependent radiative Jeffrey nanofluid flow modelling using Laplace transformInternational Journal of Numerical Methods for Heat & Fluid Flow · 2026
- Comparative study of turbulence models for hemolysis prediction and hydraulic performance in centrifugal blood pumps.Clinical biomechanics (Bristol, Avon) · 2026
- Numerical analysis for MHD Fe3O4 -Cu/blood hybrid nanofluid flow in a disk-cone geometry: Application of biomedical thermal system.Computational biology and chemistry · 2026
- Cosine-shaped stenotic arteries optimization of CPHNF flow for performance enhancement in unblocking using taguchi and response surface methodologiesChemical Physics Impact · 2026