Abstract

The PetVision project is developing an ultrafast photon detector module for next-generation time-of-flight positron emission tomography (TOF-PET). In contrast to contributions centered primarily on scanner geometry, the present work focuses on the detector technology required to translate laboratory-scale timing performance into scalable imaging modules. The approach combines optimized LYSO scintillators, low-crosstalk silicon photomultipliers, fast ASIC readout, and high-density module integration to target coincidence timing resolution below 100 ps FWHM. We summarize the detector concept, the rationale for the selected technology stack, and recent progress in component characterization and simulation. As a representative component-level result, a coincidence timing resolution of (85 ± 1) ps FWHM was measured with 3 × 3 mm 2 FBK NUV-HD-MT SiPMs coupled to 2 × 2× 3 mm 3 Fast-LGSO crystals and read out with FastIC+ electronics. These developments provide the hardware basis for future planar PET architectures with reduced scintillator usage, high flexibility, and improved accessibility.