Polymer Gel Dosimetry of Spatially-Fractionated Radiation Therapy with Kilovoltage X-rays

Olivia Moluchi, Olivia Masella, Jade Fischer, Devon Richtsmeier, Laura Dufort-Gagnon, Magdalena Bazalova-Carter

Key points

From the paper's conclusion
  • NIPAM-based polymer gel dosimeters were used to evaluate 3D spatially-fractionated radiation therapy (SFRT) delivered with kilovoltage x-rays.

Abstract

NIPAM-based polymer gel dosimeters were used to evaluate 3D spatially-fractionated radiation therapy (SFRT) delivered with kilovoltage x-rays. An optimized formulation was manufactured for optical clarity and dose response, then used for SFRT irradiations in two phantoms. A non-linear dose response of the polymer gel was identiTied with saturation occurring at ∼55Gy for a 225kVp beam of the dual-robot radiotherapy system (KOALA). Single- and multi-beam SFRT irradiations were performed and showed spatial fractionation, however with lower peak-to-valley dose ratios (PVDR) than expected. PVDRs ranged from 1.8-4.2 due to partial volume effects caused by the limited spatial resolution of the x-ray computed tomography (CT) readout.