Cherenkov imaging dosimetry of heterogenous phantom in the context of a MR-Linac

Audran Poher, Janelle Morrier, Tigran Galstian, Louis Archambault, Luc Beaulieu

Key points

From the paper's conclusion
  • This method has been compared to non-polarized Cherenkov for the dosimetry of a heterogeneous phantom irradiated by an Elekta Unity MR-Linac.

Abstract

Polarized Cherenkov imaging dosimetry presents several advantages in the context of MR-Linacs due to its independence on charge-based measurements. This method has been compared to non-polarized Cherenkov for the dosimetry of a heterogeneous phantom irradiated by an Elekta Unity MR-Linac. Polarized Cherenkov showed better agreement to Monte Carlo (MC) simulations with 96% of its pixels respecting 3% / 3 mm criteria at an interface with significant electron return effect. At the region of intersecting beams, non-polarized Cherenkov shows slight better agreement to MC with 88% of its pixels respecting 3% / 3 mm criteria (versus 82% for polarized Cherenkov) due to increased signal in this location.