Stereo Cherenkov imaging as 3D absolute dose monitoring tool for total skin electron therapy (TSET).
Yifeng Zhu, Baozhu Lu, Emily Xiong, Ching Lai, Karen Tang, Harper G Hubbeling, John P Plastaras, Michael LaRiviere,
Dept. of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, 19104. · Department of Biomedical Engineering, Dartmouth College, Hanover, NH 03755.
Key points
From the paper's conclusion- Cherenkov imaging provides a non-contact method for monitoring surface dose delivery during total skin electron therapy (TSET), where patients are treated in six postures using the Stanford standing technique.
Abstract
Cherenkov imaging provides a non-contact method for monitoring surface dose delivery during total skin electron therapy (TSET), where patients are treated in six postures using the Stanford standing technique. In this study, a multi-view Cherenkov imaging framework combined with patient-specific correction factors was investigated to improve quantitative three-dimensional (3D) surface dose estimation. Cherenkov emission was recorded during treatment using three synchronized time-gated cameras providing anterior and lateral coverage. Patient-specific anatomical models reconstructed from 3D surface scans were used to generate Monte Carlo (MC) reference dose distributions. Cherenkov images were corrected for background, camera response, and geometric distortions before conversion to pseudo-dose using in-vivo dosimetry calibration. Vertical Cherenkov intensity profiles were compared with MC dose profiles to derive patient-specific correction factors. Results showed that uncorrected Cherenkov-derived dose underestimated dose in distal and lateral body regions, while applying patient-specific corrections and multi-view imaging improved agreement with MC reference dose, supporting enhanced surface dose monitoring for TSET.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental Validation and Calibration of Cherenkov Imaging for Real-Time Surface Dosimetry in Photon Breast Therapy.Proceedings of SPIE--the International Society for Optical Engineering · 2026
- A Review of Cherenkov Imaging for Real-Time Verification in Radiation TherapyPractical Radiation Oncology · 2025
- From Radiotherapy Dataset to Cherenkov Dose Map: A Digital Linac Monte Carlo Simulation PipelinearXiv · 2026
- Monte-Carlo simulations of Cherenkov Fluence for FLASH radiotherapy dosimetryJournal of Physics: Conference Series · 2026
- Emission-associated energy deposition and Cherenkov-dose proportionality in megavoltage photon beamsarXiv · 2026
- Clinical in-vivo dosimetry for total skin electron therapy using gafchromic™ EBT4 film.Journal of applied clinical medical physics · 2026