Journal of Physics: Conference Series

Impact of photon beam quality on the radiation response of Fricke-XO-Gelatin with sorbitol

Michał Piotrowski, Piotr Maras, Marek Kozicki

Key points

From the paper's conclusion
  • The results show that the radiation response of Fricke–XO–Gelatin with sorbitol is largely independent of photon beam quality.

Abstract

In this work, a Fricke gel dosimeter with a gelatin matrix, xylenol orange, and sorbitol was investigated. Sorbitol was previously proposed by our group as an additive to improve the stability and thermo-mechanical properties of the dosimeter. Since it is important to verify that energy independence is maintained, the gel dosimeter samples were irradiated with doses in the range of 0–10 Gy using megavoltage photon beams of different qualities and subsequently analysed. Their absorbance was then measured spectrophotometrically. The results show that the radiation response of Fricke–XO–Gelatin with sorbitol is largely independent of photon beam quality.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers