Physics in medicine and biology

Detector-side neural timing calibration for TOF-PET from coincidence supervision.

Julian Thull, Yannick Kuhl, Thore Meyer, Marc Henning, Stephan Naunheim, Luis Lopes De Paiva, Vanessa Nadig, Stefan Gundacker,

Chair of Imaging and Computer Vision, RWTH Aachen University, Kopernikusstraße 16, Aachen, NRW, 52074, Germany. · Department of Physics of Molecular Imaging Systems, RWTH Aachen University, Pauwelsstraße 19, Aachen, NRW, 52074, Germany.

Computational

Key points

From the paper's conclusion
  • We evaluate the method on three heterogeneous PET systems against conventional time-skew and time-walk calibration and matched direct coincidence-level neural models.

Main results: Under a common energy-window setting, the proposed model reduced coincidence timing resolution from 321.7 to 263.0ps, 462.4 to 431.7ps, and 417.0 to 382.0ps relative to charge-based time-walk correction on the

Abstract

Machine-learning methods can model non-linear event-dependent timing effects in time-of-flight positron emission tomography (TOF-PET), but calibration measurements naturally provide timing supervision only at the coincidence-level rather than for individual detector interactions. We investigate whether coincidence measurements can be used to learn reusable detector-side timing calibration functions without individually labelled interaction times. Approach: A shared detector-conditioned neural network independently processes each single-event readout and predicts a detector-side timing correction and variance contribution. During training, two detector-side timing estimates are differenced and compared with the geometrically expected time difference obtained from point-source coincidence measurements. We evaluate the method on three heterogeneous PET systems against conventional time-skew and time-walk calibration and matched direct coincidence-level neural models. Main results: Under a common energy-window setting, the proposed model reduced coincidence timing resolution from 321.7 to 263.0ps, 462.4 to 431.7ps, and 417.0 to 382.0ps relative to charge-based time-walk correction on the three systems. In detector-pair holdout experiments, detector-side models achieved comparable or slightly better timing performance than direct coincidence-level neural models while retaining independently applicable singles-level outputs. The learned calibration generalized to unseen source positions and detector pairs, enabled recombination of independently trained scanner components after scalar timing-reference alignment, provided event-wise coincidence-timing uncertainty stratification in both full-ring systems, and remained compatible with low-precision inference after quantization-aware training. Significance: Coincidence supervision enables reusable detector-side timing corrections that improve CTR without single-event timing labels. This preserves a singles-level calibration interface while enabling machine-learning-based nonlinear timing correction, modular scanner calibration, and event-specific coincidence-timing uncertainty estimates.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers